Full name
Isabela Mirales Pires
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2019
Supervisor
Committee
Ferracioli, Renata Cristina Silveira Rodrigues (President)
Almeida, Rossana Pereira de
Assunção, Wirley Gonçalves
Rivera, Izabela Cristina Mauricio Moris
Title in Portuguese
Efeito da simulação de ciclos mastigatórios sobre diferentes tipos de componentes para implantes com conexão interna cônica
Keywords in Portuguese
Conexão cone morse
Desadaptação marginal
Implante dentário
Prótese dentária fixada por implante
Torque
Abstract in Portuguese
O objetivo deste estudo foi avaliar perda de torque e desadaptação horizontal e vertical de componentes Mini-Pilar e UCLA para implantes Cone Morse antes e após ciclagem termomecânica. Foram confeccionadas 20 próteses fixas metalocerâmicas de três elementos suportadas por dois implantes, divididas em dois grupos. Grupo 1: Mini-Pilar e Grupo 2: UCLA (n=10). Todas as próteses foram avaliadas em relação a desadaptação horizontal e vertical, utilizando MicroCT, e também em relação à perda de torque e antes e após a ciclagem termomecânica. A ciclagem foi realizada com carga de 80N e com deslizamento de 0,7mm para simular movimento excursivo, a uma frequência de 2HZ, realizando 300.000 ciclos e variação de temperatura entre 5° e 55°. As comparações estatísticas de perda de torque e desadaptação foram realizadas pelo Modelo Linear de Efeitos Mistos com nível de significância p≤0,05. A perda de torque foi analisada antes (T0) e após 300.000 ciclos (T1), onde foram comparados o parafuso protético do grupo 1 e parafuso do UCLA do grupo 2. Ambos apresentaram perda de torque antes e após a ciclagem termomecânica (p<0,05), porém, essa perda de torque foi maior para o grupo 1, antes e após ciclagem. A perda de torque do parafuso do Mini-Pilar antes e após ciclagem termomecânica também foi avaliada, a ciclagem demonstrou efeito significativo no aumento da perda de torque do parafudo de retenção do Mini-Pilar (p<0,05). Em relação a desadaptação horizontal e vertical notou-se que no grupo 1 apenas a desadaptação marginal vertical apresentou diminuição significativa estatisticamente após ciclagem termomecânica (p=0,028), por outro lado, no grupo 2 a desadaptação interna horizontal e vertical não apresentou resultados estatísticos significativos. Conclui-se que o componente UCLA teve melhor comportamento em relação a perda de torque e em relação a adaptação, apenas a adaptação marginal vertical do Mini-Pilar diminuiu significativamente após a ciclagem.
Title in English
Effect of chewing cycle simulation on different component types in implants with conical internal connection
Keywords in English
Dental implant
Implant fixed dental prosthesis
Misfit
Morse cone connection
Torque
Abstract in English
The aim of this study was to evaluate torque loss and horizontal and vertical misfit of mini-conical and UCLA abutments for Morse Cone Implants before and after thermomechanical cycling. Twenty metalloceramic fixed prostheses were made with three elements, supported by two implants and divided into two groups. Group 1: Mini conical and Group 2: UCLA (n=10). All prostheses were evaluated for horizontal and vertical misfit using MicroCT, and torque loss before and after thermomechanical cycling. Cycling was performed with 80N load and 0.7mm slide to simulate excursion movement at a frequency of 2Hz, performing 300,000 cycles and temperature range between 5° and 55°. Statistical comparisons of torque loss and misfit were performed by the Linear Mixed Effects Model with significance level p≤0.05. Torque loss was analyzed before (T0) and after 300,000 cycles (T1), where the group 1 prosthetic screw and the group 2 UCLA screw were compared.Both presented torque loss before and after thermomechanical cycling (p <0.05), but this torque loss was greater for group 1 before and after cycling. Mini conical screw torque loss before and after thermomechanical cycling was also assessed, cycling demonstrated significant effect on increased torque loss (p <0.05). Regarding horizontal and vertical misfit, it was noted that in group 1 only the vertical marginal misfit showed a statistically significant decrease (p=0.028). On the other hand, in group 2, the horizontal and vertical internal misfit did not present significant statistical results. It was concluded that the UCLA component presented better behavior regarding torque loss and fit, only the Mini conical vertical marginal misfit significantly decreased after cycling.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2022-10-03
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.