Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2021.tde-10032022-162000
Documento
Autor
Nombre completo
João Carlos Alves Prestes
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2021
Director
Tribunal
Potiens, Maria da Penha Albuquerque (Presidente)
Silva, Eric Alexandre Brito da
Santos, Lucas Rodrigues dos
Título en portugués
Caracterização de uma instalação de 60Co para calibração de dosímetros clínicos utilizados em radioterapia
Palabras clave en portugués
calibração
dosímetros clínicos
radioterapia
theratron 780c
Resumen en portugués
Este trabalho foi desenvolvido a fim de implementar as etapas de adequação e caracterização de um irradiador de 60Co, que era utilizado no tratamento para que seja utilizado em procedimentos de calibrações de dosímetros clínicos utilizados em radioterapia. Foi realizado nesse trabalho uma conformação de um sistema irradiador doado ao Laboratório de Calibração do IPEN e a caracterização do feixe e do campo de radiação, de acordo com as recomendações do guia da AIEA (TRS-469, 2009). Para a instalação do novo irradiador, foi realizada a adaptação na atual sala de calibração para acomodar o cabeçote na posição de calibração. Também fez parte o desenvolvimento de um suporte o seu posicionamento. Na sequência foi necessária a correção no sistema de colimação. Para a caracterização do feixe de radiação foram realizados os testes iniciais de uniformidade do campo de radiação determinando o tamanho e o centro do feixe de radiação, determinação das grandezas de referência kerma no ar e dose absorvida na água, variações de saída devido ao movimento da fonte, tempo de estabilização, teste de corrente de fuga, teste de repetibilidade e estabilidade a longo prazo.
Título en inglés
Characterization of a 60Co installation for calibration of clinical dosimeters used in radiotherapy
Palabras clave en inglés
calibration
dosímeters clínical
radiotherapy
theratron 780c
Resumen en inglés
This work was developed to implement the steps of adaptation and characterization of a 60Co irradiator, which was used for treatment so that it can be used in clinical dosimeter calibration procedures used in radiotherapy. In this work, an adaptation of an irradiating system donated to the IPEN Calibration Laboratory was carried out and the characterization of the beam and the radiation field, according to the recommendations of the IAEA guide (TRS-469, 2009). For the installation of the new radiator, the current calibration room was adapted to accommodate the irradiator in the calibration position. The development of a support for its positioning was also part. Then, correction in the collimation system was necessary. For the characterization of the radiation beam, initial tests of uniformity of the radiation field were performed, determining the size and center of the radiation beam, determination of the kerma reference quantities in the air and dose absorbed in the water, variations in output due to the movement of the radiation source, stabilization time, leakage current test, repeatability test and long-term stability.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2022-03-14
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.