Tese de Doutorado
DOI
https://doi.org/10.11606/T.85.2018.tde-13032018-101043
Documento
Autor
Nome completo
Iremar Alves da Silva Junior
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2017
Orientador
Banca examinadora
Potiens, Maria da Penha Albuquerque (Presidente)
Bottaro, Márcio
Caldas, Linda Viola Ehlin
Silva, Teógenes Augusto da
Vivolo, Vitor
Título em português
Desenvolvimento de uma metodologia de calibração de monitores de contaminação de superfície considerando o mapeamento da uniformidade das fontes extensas de referência
Palavras-chave em português
calibração
fator de correção
fonte extensa de referência
MCNP
monitor de contaminação de superfície
uniformidade
Resumo em português
Uma das exigências na calibração de monitores de radiação de contaminação de superfície consiste no uso de fontes extensas de referência com um valor de uniformidade superior a 90%. Entretanto, foi verificado em alguns laboratórios de calibração que suas fontes não atendiam essa exigência, o que obrigaria esses laboratórios a não utilizarem tais fontes em suas calibrações. Foi neste contexto que foi desenvolvido um estudo para propor o uso dessas fontes, desde que sejam utilizados fatores de correção, que são propostos neste trabalho, que foram avaliados e desenvolvidos usando o código de transporte MCNP como ferramenta de análise e avaliação, a partir dos dados de mapeamento da uniformidade. Também foi desenvolvido um software para calcular esses fatores de correção, utilizando o mapeamento da uniformidade das fontes como dados de entrada e um sistema automatizado para calibração de monitores de radiação de contaminação de superfície. O grande ganho com esse trabalho foi poder utilizar fontes extensas de referência, mesmo quando essas estão em desacordo com as exigências de uniformidade da norma ABNT ISO 8769:2016.
Título em inglês
Development of a calibration methodology of surface contamination monitors considering the uniformity mapping of the large area reference sources
Palavras-chave em inglês
calibration
correction factor
large-area reference source
MCNP
surface contamination monitor
uniformity
Resumo em inglês
One of the requirements in the calibration of surface contamination radiation monitors is the use of wide area reference sources with a uniformity value greater than 90%. However, it has been found, in some calibration laboratories, that their sources were not meeting this requirement, which would not allow these laboratories to use such sources in their calibration procedures. It was in this context that a study was developed to propose the use of these sources, provided that correction factors are used, which are proposed in this work, which were evaluated and developed using the MCNP radiation transport code as an evaluation tool, from the uniformity mapping data. A software was also developed to calculate these correction factors using the source uniformity mapping as an input data along with an automated system for calibration of surface contamination radiation monitors. The great gain in this work was to be able to use wide area reference sources, even when they are in disagreement with the uniformity requirements of ISO 8769:2016.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2018-04-11
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.