Tesis Doctoral
DOI
https://doi.org/10.11606/T.85.2016.tde-05122016-144851
Documento
Autor
Nombre completo
Claudio Domienikan
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2016
Director
Tribunal
Saxena, Rajendra Narain (Presidente)
Carbonari, Artur Wilson
Macchione, Eduardo Luiz Augusto
Madi Filho, Tufic
Oliveira, Jose Roberto Brandao de
Título en portugués
Desenvolvimento de um espectrômetro de correlação angular gama-gama perturbada com seis detectores de BaF2 e estudo de interações hiperfinas em composto intermetálico LaMnSi2
Palabras clave en portugués
CAP
espectrometria gama
espectrômetro gama
física do estado sólido
instrumentação nuclear
Resumen en portugués
Neste trabalho foi desenvolvido um Espectrômetro de Correlação Angular Gama- Gama Perturbada Diferencial em Tempo (CAP) constituído por seis detectores cintiladores de BaF2, para realização de medidas de interações hiperfinas (campo hiperfino magnético e gradiente de campo elétrico) em diversos materiais e propiciar estudos na área da física da matéria condensada. O espectrômetro desenvolvido possui um sistema de aquisição não convencional em comparação aos demais equipamentos destinados a medidas de CAP. Ao invés do tradicional Analisador Multicanal (MCA), este espectrômetro utiliza um sistema de aquisição de dados constituído, basicamente, por um Conversor Analógico Digital (ADC) rápido, uma placa digital (I/O) convencional e um roteador construído no laboratório de Interações Hiperfinas (LIH) do IPEN. Este versátil e eficiente sistema, controlado por um software também criado no LIH em LabVIEW, permite a geração simultânea de 30 espectros de coincidências γ - γ atrasadas, número superior em comparação aos 12 espectros do antigo espectrômetro de quatro detectores. Além de medidas de linearidade, resolução em tempo e tempo morto, o funcionamento e o desempenho do espectrômetro foram comprovados através de medidas de CAP utilizando os núcleos de prova 111In -> 111Cd e 181Hf -> 181Ta, cujos resultados são bem conhecidos da literatura. Foram feitas medidas de interação quadrupolar do 181Ta em háfnio metálico e do 111Cd em cádmio metálico, e de campo hiperfino magnético do 111Cd e do 181Ta em níquel. Os resultados destas medidas se mostraram em concordância com a literatura. Adicionalmente foram realizadas medidas inéditas de interações hiperfinas magnéticas no composto intermetálico LaMnSi2 utilizando os núcleos de prova 111Cd e 140Ce. As medidas foram realizadas na faixa de temperatura de 10 K a 400K. No caso das medidas utilizando a sonda 111In -> 111Cd, os resultados mostram uma variação do campo magnético com a temperatura que segue a função de Brillouin. Já no caso das medidas com o núcleo de prova 140La -> 140Ce, o resultado apresentou um comportamento anômalo do campo hiperfino em função de temperatura. Os resultados evidenciam uma forte hibridização da banda 4f do Ce com a banda 3d do Mn, fato verificado e estudado em trabalhos anteriores com compostos semelhantes.
Título en inglés
Development of a perturbed gama-gama angular correlation spectrometer with six BaF2 detectors and study of hiperfine interaction in the intermetallic compound LaMnSi2
Palabras clave en inglés
gamma spectrometry
nuclear instrumentation
PAC
PAC spectrometer
solid state physics
Resumen en inglés
In this work a Perturbed gama-gama Angular Correlation (PAC) spectrometer was constructed consisting of six BaF2 scintillator detectors to perform measurements of hyperfine interactions (magnetic hyperfine field and electric field gradient) in different materials to study condensed matter physics. The spectrometer developed has an unconventional acquisition system compared to other equipment for PAC measurements. Instead of a traditional multichannel analyzer (MCA), the spectrometer utilizes a data acquisition system consisting of basically a fast analog to digital converter (ADC), a conventional digital card (I/O) and a router constructed in the hyperfine interactions laboratory (LIH) of IPEN. This versatile and efficient system, controlled by software, also developed in the LIH using LabVIEW, allows simultaneous generation of 30 delayed γ - γ coincidence spectra compared to 12 spectra in the old 4 detector spectrometer in our laboratory. In addition to the tests of system linearity, time resolution and dead time, the operational performance of this spectrometer was demonstrated by PAC measurements using 111In -> 111Cd and 181Hf -> 181Ta nuclear probes, for which the results are well known from the literature. The quadrupole interaction of 181Ta in metallic hafnium, and 111Cd in metallic cadmium, and magnetic hyperfine field of 111Cd in nickel, were measured and the results are in agreement with the literature. Additionally the measurements were carried out, with the new spectrometer, to study the hyperfine field in the intermetallic compound LaMnSi2 using 111Cd and 140Ce nuclear probes. The measurements were carried out in the temperature range from 10 K to 400 K. While the temperature dependence of hyperfine field measured with 111In -> 111Cd probe, follows the Brillouin function the behaviour of the hyperfine field measured with 140La -> 140Ce is anomalous. This behavior has been explained in terms of a strong hybridization of 4f band of Ce with the 3d band of Mn, a fact verified in previous studies with similar compounds.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2016-12-15
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.