Tesis Doctoral
DOI
https://doi.org/10.11606/T.43.2004.tde-07032014-160312
Documento
Autor
Nombre completo
Fabio Henrique Palladino
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2004
Director
Tribunal
Robilotta, Cecil Chow (Presidente)
Gutierrez, Marco Antonio
Okuno, Emico
Schiabel, Homero
Tannus, Alberto
Título en portugués
Reconstrução 3D de imagens em tomografia por emissão de pósitrons com Câmaras de Cintilação
Palabras clave en portugués
Métodos iterativos
Simulações
Resumen en portugués
A Tomografia por Emissão de Pósitrons (PET) está se definindo como um dos métodos preferidos para diagnóstico e seguimento de inúmeras doenças em Oncologia, Neurologia e Cardiologia. Esta modalidade é realizada com sistemas dedicados e sistemas baseados em câmaras de cintilação, que podem ser também usados em tomografia por emissão de fótons únicos (SPECT). Neste trabalho, efetuamos uma avaliação dos fatores que favorecem a quantificação em imagens PET com câmaras de cintilação em coincidência, caracterizadas por urna menor sensibilidade em relação a sistemas dedicados. Avaliamos as condições de quantificação de imagens sob os modos 2D e 3D de aquisição, obtidas por métodos de reconstrução 2D e 3D diversos e correções associadas. Dados de aquisição foram simulados por método de Monte Carlo empregando parâmetros realistas. Objetos de interesse diversos foram modelados. Imagens foram reconstruídas pelos métodos FBP, ART, MLEM e OSEM e consideramos correções de sensibilidade, normalização de detector, espalhamento e atenuação de radiação. Estabelecemos uma metodologia de avaliação de detectabilidade e recuperação de contrastes em imagens que contemplam, a partir de dois parâmetros mensuráveis, os aspectos mais relevantes em quantificação. Análises visuais também foram consideradas. Verificamos que o modo 3D é mais adequado que 2D na recuperação de baixos contrastes no objeto com a aplicação das correções selecionadas. A detectabilidade de pequenas estruturas está limitada pelos efeitos de volume parcial e pela resolução espacial finita dos sistemas de detecção. Os métodos ART, MLEM e, em particular, OSEM com 8 subconjuntos, apresentam-se adequados para estudos quantitativos no modo 3D. Os parâmetros definidos para avaliação podem ser empregados como indicadores de condições propícias a estudos quantitativos.
Título en inglés
3D Image Reconstruction Positron Emission Tomography Scintillation Cameras
Palabras clave en inglés
Iterative methods
Simulations
Resumen en inglés
Volumetric reconstruction in gamma camera based PET imaging Positron Emission Tomography (PET) is considered as a very useful tool for diagnosing and following several diseases in Oncology, Neurology and Cardiology. Two types of systems are available for this imaging modality: the dedicated systems and those based on gamma camera technology. In this work, we assessed a number of factors affecting the quantitation of gamma camera based PET imaging, characterized by a lower sensitivity compared to those of dedicated systems. We also evaluated image quantitation conditions under 2D and 3D acquisition/reconstruction modes, for different reconstruction methods and associated corrections. Acquisition data were simulated by Monte Carla method, using realistic parameters. Several objects of interest were modelled. We reconstructed slices and volumes using FBP, ART, MLEM and OSEM and also included four corrections: detector sensitivity, detector normalization, scatter and attenuation of annihilation photons. We proposed a method to assess detectability and object contrast recovery by using two measurable parameters. Visual analysis was also considered. We found that 3D mode is more effective than 2D for low contrast recovery when the selected (J corrections are applied. Detectability of small structures is limited by partial volume effects and device finite spatial resolution. ART, MLEM and specially 8-subsets OSEM are the most adequate methods for quantitative studies in 3D mode. The parameter that we have defined may also be used as indicators of suitable conditions for quantitation in images.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
46482Palladino.pdf (15.15 Mbytes)
Fecha de Publicación
2014-03-12
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.