Nom complet
Patrícia Miranda de Faria
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 1999
Directeur
Jury
Pedrozo, Rosemary Custodio (Président)
Colacioppo, Sergio
Salgado, Paulo Eduardo de Toledo
Titre en portugais
Padronização de metodologia para a determinação de metais em amostras de cabelo utilizando um espectrômetro de massa de alta resolução acoplado a uma fonte de plasma induzido por gás de argônio
Mots-clés en portugais
Análise toxicológica
Toxicologia
Resumé en portugais
A técnica de espectrometria de massa de alta resolução com fonte de plasma induzida por argônio tem sido reconhecida pela sua sensibilidade, especificidade e capacidade para análises multielementares. Por estas características, a sua aplicação em análises de metais em amostras de cabelo começa a ser investigada em Toxicologia. Este trabalho descreve um método para determinação de cádmio, chumbo, cobalto, cromo e níquel em amostras de cabelo, utilizando um espectrômetro de alta resolução acoplado a uma fonte de plasma induzido por argônio "Element" da Finnigan MAT ®. O estudo de precisão externa mostrou coeficiente de variação de 7,79 a 17,08% e o estudo de precisão interna 0,56 a 5,84%. A exatidão do método para análises de material de referência certificado foi de 95,97% para cádmio e 99,30% para chumbo. Limites de detecção da ordem de ng/L foram obtidos para todos os elementos e linearidade na faixa de concentração de 0,5 a 50,0 µg/L com coeficientes de correlação entre concentração e intensidade do sinal de 0,9995 a 0,9999. A concentração de metais no cabelo para os seis voluntários variaram de 0,08 a 1,25 µg/g para cádmio, 1,66 a 5,39µg/g para chumbo, 0,06 a 1,12 µg/g para cobalto, 1,75 a 8,33 µg/g para o cromo e 1,08 a 2,34 µg/g para o níquel. O método foi considerado adequado para a determinação dos metais selecionados em amostras de cabelo.
Titre en anglais
Determination of trace metals in human hair by high resolution inductivel y coupled mass spectrometry
Mots-clés en anglais
Análise toxicológica
Toxicologia
Resumé en anglais
The high resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) has been recognized to be a sensit ive, specific and multi-elemental technique. According to these characteristics, it's application in hair analysis begins to be investigated in Analytical Toxicology. This paper describe a method to determine some metals (chromium, cobalt, nickel, cadmium and lead) in human hair samples, using a HR-ICP-MS equipment. Precision shows coefficient of variation from 7.79 to 17.08% (external precision) and from 0.56 to 5.84% (internal precision). Recovery using certified sample was 95.97% for cadmium and 99.30% for lead. Limits of detection were obtained at ng L-1 levels for all elements. Results were in a good agreement with correlation coefficients, between 0.9995 to 0.9999. Present data indicate that HR-ICP-MS is applicable to a rapid and accurate multi-element determination, virtually free from spectral interference. The concentrations of the metals in hair samples of six volunteers varied from 0.08 to 1,25 µg g-1 for cadmium; 1.66 to 5.39 µg g-1 for lead; 0.06 to 1.12 µg g-1 for cobalt; 1. 75 to 8.33 µg g-1 for chromium and 1.08 to 2 .34 µg g-1 for nickel. The methodolo9y was consider adequate to determine metal concentrations in hair.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2022-08-24
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.