• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.85.2014.tde-29092015-083645
Documento
Autor
Nombre completo
Luís Cláudio Aranha
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2014
Director
Tribunal
Rossi, Jesualdo Luiz (Presidente)
Bonachela, Wellington Cardoso
Braga, Francisco Jose Correa
Casemiro, Luciana Assirati
Mucsi, Cristiano Stefano
Título en portugués
Estudo dos parâmetros de estabilidade de travamento entre implante dentário e componente protético através do sistema cone Morse
Palabras clave en portugués
cone Morse
contaminação
fadiga
implante
microtorque
Resumen en portugués
Na implantodontia visa-se dois grandes resultados para os desdentados parciais e totais, função mastigatória e estética. Ambos têm o mesmo grau de importância, para tanto se deve pensar no que se tem de mais seguro e prático nos dias de hoje. O cone Morse, na implantodontia oral, é um sistema de estabilidade entre o implante dentário e o parafuso intraósseo de sustentação. Para seu correto uso se faz necessário uma série de análise de suas características e condições. No presente trabalho focou-se em desenvolver protocolo de validação de torque, além de analisar parâmetros de acabamento, deformação e estabilidade précarga e pós-carga dentro de protocolos já recomendados. Avaliou-se sua condição antes e após os efeitos de carga mastigatória simulada por ensaio de fadiga. Concluiu-se que o uso do cone Morse dentro dos protocolos indicados deve ser revisto quanto a padrões similares de torque nos pilares sólidos e indexados, cada um possui limites e indicações diferentes para se manter estável. Concluiu-se também que não é um sistema totalmente livre de trânsito de bactérias e que esse trânsito aumenta de forma proporcional ao uso e força mastigatória, indicando que melhorias no acabamento entre as peças devem ser implementadas e/ou adição de material que afaste crescimento bacteriano priorizando a saúde do periodonto e do tecido ósseo de sustentação mantendo o sistema seguro em longevidade.
Título en inglés
Study on the stability parameters of locking between dental implants and prosthetic component system through the Morse taper
Palabras clave en inglés
contamination
fatigue
implant
microtorque
Morse taper
Resumen en inglés
In implantology it is aimed two major outcomes for the partial and total toothless persons, mastigatory function and aesthetics. Both have the same degree of importance, therefore it must be thought about what is safer and more practical these days. Morse taper in oral implantology is a system of stability between the dental implant and the intraosseous screw support. For its correct use a series of analysis of its features and conditions is required. This work was focused on developing torque validation protocol, and analyzing surface finishing parameters, deformation and stability as before load and after load in recommended protocols. The implants condition before and after the effects of habitual masticatory load was simulated by fatigue testing. It was concluded that the Morse taper use within the specified protocols should be reviewed for similar patterns of torque in solids and indexed pillars, each has different limits and signs to remain stable. It was also concluded that it is not a traffic totally free system of bacteria and that this influence increases in proportion to the use and masticatory forces, improvements in the surface finish between parts and / or addition of material that hinders bacterial growth is needed if it is prioritized the health of the periodontium and the bone tissue keeping the system safe in the long therm.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
2014AranhaEstudo.pdf (3.42 Mbytes)
Fecha de Publicación
2015-10-09
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.