Estudos dosimétricos utilizando um sistema 3D para controle de qualidade em radioterapia
Palabras clave en portugués
ArcCHECK
controle de qualidade
dosimetria
radioterapia
simulação Monte Carlo
Resumen en portugués
Nos últimos anos, foi desenvolvido e comercializado um moderno sistema dosimétrico 3D composto por 1386 diodos arranjados de forma helicoidal, chamado ArcCHECK, para controle de qualidade de técnicas que envolvem modulação dos feixes na radioterapia. O escopo deste trabalho é contribuir para estudos dosimétricos para o controle de qualidade, através da análise e da compreensão de medidas experimentais realizadas com o ArcCHECK e de simulações computacionais no código MCNP, baseado no Método Monte Carlo. Foram observadas as respostas do sistema para irradiações simples e estáticas, comparadas aos resultados de dois modelos computacionais desenvolvidos para simulações, além disso o método de interpolação e a sensibilidade de detecção do sistema foram testados. Foram encontradas boas reprodutibilidade, repetitividade, linearidade com a dose e com a taxa de dose e sensibilidade de detecção, embora alguns efeitos inesperados tenham sido observados principalmente devido à representação da aquisição helicoidal da dose em um mapa de distribuição 2D. O modelo computacional proposto de forma mais simples, assim como o algoritmo desenvolvido para interpolação dos dados, auxiliou na compreensão dos efeitos do ArcCHECK. O modelo mais detalhado proposto foi validado e pode ser usado como referência para diferentes simulações futuras.
Título en inglés
Dosimetric studies using a 3D system for quality assurance in radiation therapy
Palabras clave en inglés
ArcCHECK
dosimetry
Monte Carlo simulation
quality assurance
radiation therapy
Resumen en inglés
In the last years, a modern 3D dosimetric system consisting of 1386 helically arranged diodes called ArcCHECK has been developed and marketed for quality assurance of techniques involving beam modulation in radiotherapy. The scope of this work is to contribute to dosimetric studies for quality assurance by analysing and understanding experimental measurements performed with ArcCHECK and computational simulations in MCNP code, based on Monte Carlo Method. The system responses for simple and static irradiations were observed, compared to the results of two computational models developed for simulations, as well as the interpolation method and the detection sensitivity of the system were tested. Good reproducibility, repeatability, dose and dose rate linearity and detection sensitivity were found, although some unexpected effects were observed mainly due to the representation of helical dose acquisition on a 2D distribution map. The simplest proposed computational model, as well as the algorithm developed for data interpolation, helped to understand the effects of ArcCHECK. The detailed model was validated and it can be used as a reference for different simulations in the future.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
2019MazerEstudos.pdf (2.96 Mbytes)
Fecha de Publicación
2020-02-07
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.