• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.85.2014.tde-31032014-145823
Documento
Autor
Nombre completo
Patrícia de Lara Antonio
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2013
Director
Tribunal
Caldas, Linda Viola Ehlin (Presidente)
Bueno, Carmen Cecília
Nersissian, Denise Yanikian
Pinto, Teresa Cristina Nathan Outeiro
Vivolo, Vitor
Título en portugués
Estabelecimento de padronização primária e de métodos relativos com o uso de técnicas luminescentes em dosimetria da radiação beta
Palabras clave en portugués
câmara de extrapolação
dosimetria
luminescência opticamente estimulada
radiação beta
termoluminescência
Resumen en portugués
Devido à inexistência de um laboratório de padronização primária na América Latina, para calibração e dosimetria de fontes de radiação beta, neste trabalho uma câmara de extrapolação Böhm foi caracterizada e estabelecida como um sistema padrão primário no Laboratório de Calibração de Instrumentos (LCI) do IPEN. Como objetivo principal deste trabalho, a câmara de extrapolação Böhm foi caracterizada com relação à sua resposta em feixes padrões secundários de radiação beta de fontes de 90Sr+90Y, com duas janelas de entrada diferentes: Mylar aluminizado e Hostaphan grafitado, e com dois métodos de determinação de taxa de dose absorvida, para verificação do seu desempenho com cada material. Embora se trate do instrumento mais adequado para a calibração de detectores e fontes de radiação beta, a câmara de extrapolação deve ser utilizada apenas em laboratórios de calibração e não em programas de controle de qualidade, pois ela tem 7 kg e é de alto custo. Assim, é importante ter sistemas considerados padrões de trabalho e métodos alternativos para substituir a câmara de extrapolação em medições realizadas em campo, como em hospitais ou clínicas. Por este motivo, foram caracterizadas três câmaras de ionização de placas paralelas, com o objetivo de se verificar a possibilidade de seu uso como sistemas padrões de trabalho em programas de controle de qualidade, e também foram estudados materiais dosimétricos diferentes, em campos de radiação beta de 90Sr+90Y, utilizando-se os fenômenos da termoluminescência (TL) e da luminescência opticamente estimulada (LOE), como métodos alternativos na calibração de fontes de radiação beta. Inclui-se aqui o desenvolvimento de um sistema postal dosimétrico para a calibração de aplicadores clínicos de 90Sr+90Y. Os resultados obtidos com a câmara de extrapolação Böhm permitiram o seu estabelecimento como sistema padrão primário na calibração de fontes de 90Sr+90Y, pois a câmara com as duas janelas de entrada diferentes permitiram a determinação de taxas de dose absorvida no ar e no tecido, à profundidade nula e a 0,07 mm. A caracterização das três câmaras de ionização de placas paralelas mostraram que elas podem ser usadas em programas de controle de qualidade de fontes de radiação beta, pois suas respostas se apresentaram adequadas para esta finalidade. Os materiais dosimétricos testados com relação às suas respostas TL e LOE também se mostraram adequados para uso em campos de radiação beta, pois suas respostas foram satisfatórias. O sistema dosimétrico, utilizando dosímetros TL, foi testado em hospitais e, posteriormente, foi enviado na forma postal a um laboratório, contendo detectores TL e LOE, e o resultado da calibração dos aplicadores clínicos foi adequado.
Título en inglés
Establishment of primary standardization and relative methods with the use of luminescent techniques in beta radiation dosimetry
Palabras clave en inglés
beta radiation
dosimetry
extrapolation chamber
optically stimulated luminescence
thermoluminescence
Resumen en inglés
Due to the lack of a primary standard laboratory at the Latin America, for the calibration and dosimetry of beta radiation sources, it was decided to establish the Böhm extrapolation chamber at the Calibration Laboratory (LCI) at IPEN as a primary standard. As a main objective of this work, the Böhm extrapolation chamber was characterized in relation to its response in beta radiation secondary standard beams of 90Sr+90Y sources, using two different entrance windows: aluminized Mylar and graphited Hostaphan, and using two methods of determination of absorbed dose rates, to verify its performance in each material. Although it is the most adequate instrument for the calibration of beta radiation detectors and sources, the extrapolation chamber should only be used at calibration laboratories, and not in quality control programs, because it weights 7 kg, and it is expensive. Thus, it is important to have systems considered work standards and alternative methods to replace the extrapolation chamber in measurements performed outside laboratories, such as in hospitals and clinics. For this reason, three parallel plate ionization chambers were characterized, with the objective to verify the possibility of their use as work standard systems, and different dosimetric materials were also studied, in beta radiation fields, using the thermoluminescence (TL) and optically stimulated luminescence (OSL) phenomena, as alternative methods for the calibration of beta radiation sources. The development of a dosimetric postal system for the calibration of 90Sr+90Y clinical applicators was included in this objective. The results obtained using the Böhm extrapolation chamber allowed its establishment as a primary standard system at the calibration of 90Sr+90Y sources, because it was possible to determine the absorbed dose rates in air and in the tissue, at null depth and 0.07 mm using the chamber with the two different entrance windows. The characterization of the three parallel plate ionization chambers showed that they can be used in quality control programs of beta radiation sources, because their responses were adequate to this purpose. The dosimetric materials were tested in relation to their TL and OSL response, and their use showed also to be adequate in beta radiation fields. The dosimetric postal system, using TL dosimeters, was tested in hospitals, and afterwards, it was sent to a laboratory, by conventional mail, with TL and OSL detectors, and the results of the calibration of the clinical applicators was adequate.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-04-07
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.