Master's Dissertation
DOI
https://doi.org/10.11606/D.85.2014.tde-16062014-155542
Document
Author
Full name
Tallyson Sarmento Alvarenga
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2014
Supervisor
Committee
Caldas, Linda Viola Ehlin (President)
Pinto, Teresa Cristina Nathan Outeiro
Vivolo, Vitor
Title in Portuguese
Instalação e caracterização básica de um laboratório para testes de monitores portáteis com radiação de nêutrons
Keywords in Portuguese
laboratório
monitores
nêutrons
Abstract in Portuguese
Foi realizada a implantação de um laboratório para testes de detectores com radiação de nêutrons. Uma sala foi reformada e adaptada para ser o Laboratório de TestescomNêutrons(LTN) no subsolo do prédio do Laboratório de Calibração de Instrumentos do IPEN. Foi criada uma sala de controle do sistema de exposição e posicionamento da fonte e para o operador responsável pelos procedimentos de testes dos instrumentos. Foi desenvolvida uma blindagem adequada para a fonte de 241Am(Be), previamente calibrada no Laboratório Nacional de Metrologia das Radiações Ionizantes (LNMRI), num sistema padrão primário. Um arranjo de irradiação automatizado foi implantado, possibilitando a variação da distância fonte-detector por meio de um comando à distância por um painel eletrônico localizado na sala de controle. Os resultados obtidos do levantamento radiométrico permitiram avaliar as condições existentes no LTN e a sua classificação. Foi realizado um estudo de estabilidade a curto e a médio prazo da resposta de vários detectores de nêutrons, que apresentaram resultados dentro do limite recomendado pela norma internacional. A influência da radiação espalhada foi determinada por duas técnicas: Método de Monte Carlo e método experimental, com a utilização de um cone de sombra e um detector de radiação de nêutrons portátil calibrado no LNMRI. Alguns monitores portáteis foram testados no arranjo estabelecido. Dos resultados obtidos por meio da simulação foi possível concluir que a estrutura do LTN está em concordância com os resultados obtidos em outros laboratórios que oferecem o mesmoserviço.
Title in English
Installation and basic characterization of a laboratory for testing of portable monitors with neutron radiation
Keywords in English
laboratory
monitors
neutron
Abstract in English
A laboratory for testing of detectors with neutron radiation was established. A room was renovated and adapted for the Testing Laboratory with Neutrons (LTN). A control room was created for the exposure and positioning of the source system and for localization of the operator responsible for the testing procedures of the instruments. An adequate shielding was developed for the 241Am(Be) source, previously calibrated at the National Metrology Laboratory of Ionizing Radiation (LNMRI), using a primary standard system. The set-up of the automated irradiation was established for the variation of the source-detector distance through a remote control by an electronic panel located in the control room. A radiometric survey was undertaken to evaluate the conditions at LTN and its classification. A study of the short and medium-term stability of the response of various detectors was realized, and the results were within the international recommended limits. The influence of the scattered radiation was determined by two techniques: Monte Carlo method and experimentally using a shadow cone and a portable radiation detector calibrated at LNMRI. Some portable monitors were tested at the established set-up. From the results obtained through the simulation, It was shown that the LTN structure is in agreement with the results obtained in other laboratories that offer the same service.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-08-14
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.