Título em português
Avaliação de dispositivos de proteção individual utilizados em radiologia diagnóstica
Palavras-chave em português
avental plumbífero
blindagem
radioproteção
raios X
Resumo em português
Neste trabalho foram avaliados aventais de proteção e blindagens de ovário de cinco fabricantes segundo as normas NBR / IEC 61331-1 e NBR / IEC 61331-3 de 2004. Para a determinação da espessura equivalente de atenuação foram aplicadas três metodologias diferentes (interpolação linear, modelo de Archer e somatório de exponenciais), uma vez que a norma não indica como este valor deve ser obtido. Além disso, foram estimadas as incertezas associadas a cada método e a influência do número de medições na incerteza combinada. As características de confecção avaliadas foram: documento acompanhante, projeto, materiais, dimensões e marcação da etiqueta. Para esta avaliação foi elaborado um formulário que contemplava todas as exigências da NBR / IEC 61331-3. Os resultados mostraram uma grande diferença entre os valores de espessura equivalente de atenuação declarados e medidos. A comparação dos resultados das três metodologias aplicadas mostrou que há pouca variação entre os valores obtidos e entre as incertezas associadas nas diferentes metodologias. Foi possível observar também que o número de medições não contribui significativamente para o aumento da incerteza nas três metodologias. A metodologia que melhor se adequou à rotina do laboratório foi a metodologia da interpolação linear com cinco medições da taxa de kerma no ar. A discrepância entre os resultados apresentados neste trabalho e as exigências das normas estudadas mostra a necessidade de tornar compulsória a certificação dos dispositivos de proteção, contribuindo assim para o aumento da proteção radiológica do usuário.
Título em inglês
Evaluation of personal protective devices used in diagnostic radiology
Palavras-chave em inglês
attenuation
data covariances
experimental data
ionization chambers
kerma
protective clothing
radiation protection
radiation quality
safety standards
thickness
X-ray equipment
Resumo em inglês
In this work protective devices of five manufacturers were evaluated according to the NBR/IEC 61331-1 and NBR/IEC 61331-3 standards. Three different methodologies (linear interpolation, Archer model and sum of exponentials) were applied for the determination of the attenuation equivalent, since the standard does not indicate how it must be determined. Moreover, the uncertainties associated to each method, and the influence of the number of measurements in the combined uncertainty were estimated. The evaluated manufacture characteristics were: accompanying document, design, materials, dimensions and label marking. For this evaluation a check list about the requirements of the NBR/IEC 61331-3 standard was elaborated. The results showed a great difference between nominal and measured attenuation equivalent values. The comparison of the results using the three methodologies showed small variations among the obtained values and among the associated uncertainties in the different methodologies. It was possible to observe that the number of measurements does not contribute significantly for the increase of the uncertainty in all three methodologies. The best methodology for the laboratory routine is the linear interpolation methodology, with five measurements for each air kerma rate value. The discrepancy between the results obtained in this work and the requirements of the applied standards show the need to adopt a compulsory certification process for protective devices, thus contributing for the increase of the radiation protection of the users.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2012-05-30
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.