Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2014.tde-15102014-105653
Document
Auteur
Nom complet
Felipe Favaro Capeleti
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2014
Directeur
Jury
Rodrigues, Letícia Lucente Campos (Président)
Galante, Ana Maria Sisti
Nersissian, Denise Yanikian
Titre en portugais
Dosimetria em tomografia computadorizada e avaliação do perfil de dose empregando dosímetro Fricke gel e a técnica de imageamento por ressonância magnética
Mots-clés en portugais
Fricke Gel
perfil de dose
ressonância magnética
tomografia
Resumé en portugais
Neste trabalho foi desenvolvido um novo método para a determinação das doses absorvidas nos exames de tomografia computadorizada, utilizando a dosimetria com Fricke gel desenvolvido no IPEN. Foram determinadas as doses absorvidas por diferentes metodologias de análise, como a espectrometria de absorção óptica, câmaras de ionização e imageamento por ressonância magnética. Foram realizados testes de limite mínimo de sensibilidade da solução de Fricke gel, repetitividade de sinal da solução de Fricke gel e dos equipamentos de tomografia computadorizada, sensibilidade de detecção, entre outros. Foram utilizados diferentes equipamentos de tomografia computadorizada com múltiplos detectores. A solução Fricke gel apresentou repetitividade melhor que ± 5,5% utilizando a técnica de espectrofotometria de absorção óptica e os equipamentos de tomografia computadorizada apresentaram repetitividade melhor que ± 0,2%. Foi desenvolvido um objeto simulador de crânio preenchido com a solução Fricke gel para a determinação da colimação total do equipamento e a distribuição de dose utilizando a técnica de imageamento tridimensional por ressonância magnética. A solução Fricke gel apresenta um preparo fácil e relativamente rápido, porém é necessária atenção para não haver contaminação e perder a solução. Com os resultados obtidos foi confirmada a aplicação deste tipo de dosimetria para os equipamentos de tomografia computadorizada.
Titre en anglais
Dosimetry for computed tomography and evaluation the dose profile using Fricke Gel dosimetry and magnetic ressonance imaging
Mots-clés en anglais
dose profile
Fricke Gel
Magnetic resonance
tomography
Resumé en anglais
In this work it was determined a new method for the determination of absorbed doses in Computed Tomography (CT) examinations using Fricke gel dosimetry developed at IPEN. Absorbed doses were determined by different methods of analysis, such as optical absorption spectrometry, ionization chambers and magnetic resonance imaging. Lower limit of sensitivity of the Fricke gel solution, the solution repeatability signal Fricke gel and CT equipment, detection sensitivity, among other tests were performed. Different equipment of computed tomography with multiple detectors were used. The Fricke gel solution showed better repeatability than 5.5% using the technique of optical absorption spectrophotometry and computed tomography equipment showed repeatability better than ± 0.2%. It was developed a skull phantom which was filled with Fricke gel solution for determining the total collimation of the equipment and the dose distribution using the technique of three-dimensional magnetic resonance imaging. The Fricke gel solution features an easy and relatively quick preparation, but it is necessary to be careful not to contaminate and lose the solution. With the results it was confirmed the application of this type of dosimetry for computed tomography equipment.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2014-11-07
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.