Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2014.tde-15102014-105653
Documento
Autor
Nombre completo
Felipe Favaro Capeleti
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2014
Director
Tribunal
Rodrigues, Letícia Lucente Campos (Presidente)
Galante, Ana Maria Sisti
Nersissian, Denise Yanikian
Título en portugués
Dosimetria em tomografia computadorizada e avaliação do perfil de dose empregando dosímetro Fricke gel e a técnica de imageamento por ressonância magnética
Palabras clave en portugués
Fricke Gel
perfil de dose
ressonância magnética
tomografia
Resumen en portugués
Neste trabalho foi desenvolvido um novo método para a determinação das doses absorvidas nos exames de tomografia computadorizada, utilizando a dosimetria com Fricke gel desenvolvido no IPEN. Foram determinadas as doses absorvidas por diferentes metodologias de análise, como a espectrometria de absorção óptica, câmaras de ionização e imageamento por ressonância magnética. Foram realizados testes de limite mínimo de sensibilidade da solução de Fricke gel, repetitividade de sinal da solução de Fricke gel e dos equipamentos de tomografia computadorizada, sensibilidade de detecção, entre outros. Foram utilizados diferentes equipamentos de tomografia computadorizada com múltiplos detectores. A solução Fricke gel apresentou repetitividade melhor que ± 5,5% utilizando a técnica de espectrofotometria de absorção óptica e os equipamentos de tomografia computadorizada apresentaram repetitividade melhor que ± 0,2%. Foi desenvolvido um objeto simulador de crânio preenchido com a solução Fricke gel para a determinação da colimação total do equipamento e a distribuição de dose utilizando a técnica de imageamento tridimensional por ressonância magnética. A solução Fricke gel apresenta um preparo fácil e relativamente rápido, porém é necessária atenção para não haver contaminação e perder a solução. Com os resultados obtidos foi confirmada a aplicação deste tipo de dosimetria para os equipamentos de tomografia computadorizada.
Título en inglés
Dosimetry for computed tomography and evaluation the dose profile using Fricke Gel dosimetry and magnetic ressonance imaging
Palabras clave en inglés
dose profile
Fricke Gel
Magnetic resonance
tomography
Resumen en inglés
In this work it was determined a new method for the determination of absorbed doses in Computed Tomography (CT) examinations using Fricke gel dosimetry developed at IPEN. Absorbed doses were determined by different methods of analysis, such as optical absorption spectrometry, ionization chambers and magnetic resonance imaging. Lower limit of sensitivity of the Fricke gel solution, the solution repeatability signal Fricke gel and CT equipment, detection sensitivity, among other tests were performed. Different equipment of computed tomography with multiple detectors were used. The Fricke gel solution showed better repeatability than 5.5% using the technique of optical absorption spectrophotometry and computed tomography equipment showed repeatability better than ± 0.2%. It was developed a skull phantom which was filled with Fricke gel solution for determining the total collimation of the equipment and the dose distribution using the technique of three-dimensional magnetic resonance imaging. The Fricke gel solution features an easy and relatively quick preparation, but it is necessary to be careful not to contaminate and lose the solution. With the results it was confirmed the application of this type of dosimetry for computed tomography equipment.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-11-07
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.