• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2017.tde-12122016-101857
Document
Auteur
Nom complet
Ladyjane Pereira Fontes
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2016
Directeur
Jury
Potiens, Maria da Penha Albuquerque (Président)
Maia, Ana Figueiredo
Nogueira, Maria do Socorro
Titre en portugais
Desenvolvimento de um Guia orientativo para dosimetria em Tomografia Computadorizada
Mots-clés en portugais
calibração
dosimetria
sistema Tandem
tomografia computadorizada
Resumé en portugais
Devido as frequentes dúvidas dos usuários das câmaras de ionização do tipo lápis calibradas no Laboratório de Calibração de Instrumentos do IPEN (LCI IPEN), em como aplicar corretamente os fatores indicados em seus certificados de calibração, foi elaborado um guia orientativo para dosimetria em tomografia computadorizada. O guia orientativo inclui o conhecimento prévio da Camada Semirredutora (CSR), já que é necessário conhecer a energia efetiva do feixe para aplicação do fator de correção para qualidade (kq). A avaliação da CSR em tomógrafos torna-se uma difícil tarefa devido à geometria do sistema e por essa razão foi realizado um levantamento das metodologias existentes para a determinação da CSR em feixes clínicos de Tomografia Computadorizada (TC), levando em consideração fatores técnicos, práticos e econômicos. Neste trabalho, optou-se em testar um Sistema Tandem composto por capas absorvedoras confeccionado na oficina do IPEN, baseado em estudos preliminares devido ao baixo custo e boa reposta. O Sistema Tandem é composto por 5 capas absorvedoras cilíndricas de 1 mm, 3 mm, 5 mm, 7 mm, e 10 mm de alumínio e por 3 capas absorvedoras cilíndricas de 15 mm, 25 mm e 35 mm de acrílico (PMMA) acopladas à câmara de ionização do tipo lápis comercial amplamente utilizada em testes de controle de qualidade na dosimetria em feixes clínicos de Tomografia Computadorizada. Através das curvas Tandem, foi possível avaliar os valores de CSR e, através da curva de calibração da câmara de ionização do tipo lápis, encontrar o Kq adequado ao feixe. O Guia elaborado traz informações de como construir a curva de calibração em função da CSR, para encontrar o Kq, bem como informações para construção da curva Tandem, para encontrar valores próximos de CSR.
Titre en anglais
Development of a guidance guide for dosimetry in computed tomography
Mots-clés en anglais
calibration
computed tomography
dosimetry
Tandem system
Resumé en anglais
Due to frequent questions from users of ionization chambers pencil type calibrated in the Instrument Calibration Laboratory of the Institute of Energy and Nuclear Research (LCI - IPEN), on how to properly apply the factors indicated in their calibration certificates, a guide was prepared guidance for dosimetry in computed tomography. The guide includes guidance prior knowledge of half value layer (HVL), as it is necessary to know the effective beam energy for application quality for correction factor (kq). The evaluation of HVL in TC scanners becomes a difficult task due to system geometry and therefore a survey was conducted of existing methodologies for the determination of HVL in clinical beams Computed Tomography, taking into account technical, practical and economic factors. In this work it was decided to test a Tandem System consists of absorbing covers made in the workshop of IPEN, based on preliminary studies due to low cost and good response. The Tandem system consists of five cylindrical absorbing layers of 1mm, 3mm, 5mm, 7mm and 10mm aluminum and 3 cylindrical absorbing covers 15mm, 25mm and acrylic 35mm (PMMA) coupled to the ionization chamber of commercial pencil type widely used in quality control tests in dosimetry in clinical beams Computed tomography. Through Tandem curves it was possible to assess HVL values and from the standard curve pencil-type ionization chamber, Kq find the appropriate beam. The elaborate Guide provides information on how to build the calibration curve on the basis of CSR, to find the Kq and information for construction Tandem curve, to find values close to CSR.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2017-03-08
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.