Doctoral Thesis
DOI
https://doi.org/10.11606/T.85.2017.tde-11082017-090607
Document
Author
Full name
Lucas Rodrigues dos Santos
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2017
Supervisor
Committee
Potiens, Maria da Penha Albuquerque (President)
Caldas, Linda Viola Ehlin
Pinto, Teresa Cristina Nathan Outeiro
Santos, William de Souza
Vivolo, Vitor
Title in Portuguese
Desenvolvimento de um protocolo de calibração utilizando espectrometria e simulação matemática, em feixes padrões de raios x
Keywords in Portuguese
coeficiente de calibração
espectrometria
função resposta
kerma
simulação pelo método de Monte Carlo
Abstract in Portuguese
A calibração, por definição, é o processo pelo qual se estabelece uma relação entre valores de medição de um padrão, com as suas respectivas incertezas, e as indicações com as incertezas associadas do instrumento de medição a ser calibrado. Um protocolo de calibração descreve a metodologia a ser aplicada em um processo de calibração. O método escolhido para a obtenção deste protocolo foi o da espectrometria de feixe de raios X associada à simulação pelo método de Monte Carlo, fundamentado no fato de que ambos são considerados métodos absolutos na determinação de parâmetros de feixes de radiação. Neste trabalho foi utilizado o método de Monte Carlo utilizado para obter a função resposta do detector utilizada para a correção dos espectros obtidos do feixe primário de radiação X; deste modo foram calculadas as taxas de kerma destes feixes e comparadas aos valores obtidos com as câmaras de ionização padrão secundário do Laboratório de Calibração de Instrumentos do IPEN (LCI/IPEN). Foram obtidos os coeficientes de calibração para o sistema padrão com diferenças em relação ao fornecido pelo laboratório primário entre 1,3% e 15,3%. Os resultados obtidos indicaram a viabilidade do estabelecimento deste protocolo de calibração utilizando a espectrometria como padrão de referência, com incertezas relativas de 0,62% para k=1. As incertezas associadas ao método proposto foram satisfatórias, para um laboratório padrão secundário e comparáveis a um laboratório primário.
Title in English
Development of a calibration protocol using spectrometry and mathematical simulation, in x ray standard beams
Keywords in English
coefficient of calibration
kerma
response function
simulation by the Monte Carlo method
spectrometry
Abstract in English
Calibration, by definition, is the process that establishes a relation between the measurement values of a standard, with their respective uncertainties, and the indications with the associated uncertainties of an instrument to be calibrated. A calibration protocol describes the methodology to be applied during the calibration process. The chosen method to obtain this protocol was the use of the X-ray beam spectrometry associated with Monte Carlo mathematical simulation, based on the fact that both are considered primary methods for x radiation beams parameters determination. The work consists in obtaining the detector response function using the Monte Carlo method for the correction of the X radiation primary beam spectra obtained. After that, the kerma rates values were calculated and compared to those obtained with the secondary standard ionization chambers of the Calibration Laboratory of IPEN. The calibration coefficients for the standard ionization chambers were found with differences from that provided by the primary laboratory between 1.3% and 15.31%. The results indicated the feasibility of establishing a protocol for the calibration of instruments using spectrometry as a reference standard, with relative uncertainties of 0.62% for k = 1. The uncertainties associated with the proposed method were satisfactory for a secondary standard laboratory and comparable to a primary laboratory.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2017-09-08
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.