Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2010.tde-01062011-122808
Document
Auteur
Nom complet
Daniel Menezes Dias
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2010
Directeur
Jury
Potiens, Maria da Penha Albuquerque (Président)
Nersissian, Denise Yanikian
Vivolo, Vitor
Titre en portugais
Estabelecimento de um novo método de calibração de câmaras de ionização tipo lápis para dosimetria em feixes de tomografia computadorizada
Mots-clés en portugais
câmara de ionização tipo lápis
dosimetria em tomografia computadorizada
metrologia em CT
radiodiagnóstico
Resumé en portugais
As câmaras de ionização tipo lápis são utilizadas para dosimetria de feixes em equipamentos de tomografia computadorizada (CT). Neste estudo foi estabelecida uma nova metodologia de calibração destas câmaras, visando a adequação do Laboratório de Calibração de Instrumentos do Instituto de Pesquisas Energéticas e Nucleares (LCI) aos atuais padrões metrológicos internacionais que tratam dos procedimentos específicos para calibração das câmaras utilizadas em CT. Primeiramente partiu-se da implantação das novas qualidades de radiação padrão RQT, que simulam feixes não atenuados usados para aplicações em CT, de acordo com a IEC 61267 da International Electrotechnical Commission (IEC). Após o estabelecimento dos campos padrões de radiação, foi implantada a metodologia de calibração específica para câmaras lápis segundo o Technical Report Series No. 457, da International Atomic Energy Agency (IAEA), que descreve as particularidades do procedimento a ser seguido pelos Laboratórios de Dosimetria Padrão Secundário (SSDLs) em termos de colimação e posicionamento em relação ao feixe de radiação. Inicialmente, foram realizadas medições de PPV (kV) e determinação das filtrações adicionais de cobre por meio dos valores de camada semi-redutora (CSR) recomendados pela norma IEC 61267, para que as qualidades de radiação de referência: RQT 8, RQT 9 e RQT 10 fossem implantadas. Para os filtros adicionais, foram utilizados alumínio e cobre de alta pureza (maior ou igual a 99,9%). Uma metodologia alternativa para determinar as filtrações adicionais também foi utilizada, através da qual se encontrou diretamente as RQTs em espessuras de filtros de cobre equivalente ao conjunto RQR (Al) + Filtração Adicional (Cu), sendo esta uma opção quando não se têm implantadas as RQRs. Com a implantação desta nova metodologia de calibração das câmaras de ionização tipo lápis, o LCI está pronto para calibrar estes instrumentos de acordo com as recentes recomendações internacionais. Permitindo assim uma melhoria na rastreabilidade das calibrações, como também na qualidade dos serviços metrológicos que o IPEN oferece para todo o País.
Titre en anglais
Establishment of a new calibration method of pencil ionization chamber for dosimetry in computed tomography
Mots-clés en anglais
dosimetry in computed tomography
pencil ionization chamber
Resumé en anglais
Pencil ionization chambers are used for beam dosimetry in computed tomography equipment (CT). In this study, a new calibration methodology was established, in order to make the Calibration Laboratory of Instituto de Pesquisas Energéticas e Nucleares (LCI) suitable to international metrological standards, dealing with specific procedures for calibration of these chambers used in CT. Firstly, the setup for the new RQT radiation qualities was mounted, in agreement with IEC61267 from the International Electrotechnical Commission (IEC). After the establishment of these radiation qualities, a specific calibration methodology for pencil ionization chambers was set, according to Technical Report Series No. 457, from the International Atomic Energy Agency (IAEA), which describes particularities of the procedure to be followed by the Secondary Standard Dosimetry Laboratories (SSDLs), concerning to collimation and positioning related to the radiation beam. Initially, PPV (kV) measurements and the determination of copper additional filtrations were carried out, measuring the half value layers (HVL) recommended by the IEC 61267 standard, after that the RQT 8, RQT 9 and RQT 10 radiation quality references were established. For additional filters, aluminum and copper of high purity (around 99.9%) were used. RQTs in thickness of copper filters equivalent to the set RQR (Al) + Additional Filtration (Cu) was directly found by an alternative methodology used to determine additional filtrations, which is a good option when RQRs have not the possibility of be setting up. With the establishment of this new methodology for the ionization pencil chambers calibration, the LCI is ready to calibrate these instruments according to the most recent international standards. Therefore, an improvement in calibration traceability, as well as in metrological services offered by IPEN to all Brazil is achieved.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2011-06-21
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.