Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.76.1995.tde-30012008-083030
Document
Auteur
Nom complet
Edson Luiz Géa Vidoto
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Carlos, 1995
Directeur
Jury
Panepucci, Horácio Carlos (Président)
Guimarães Filho, Alberto Passos
Slaets, Jan Frans Willem
Titre en portugais
Projeto de transdutores e otimização do sistema de recepção do tomógrafo de RMN de campo magnético ultrabaixo
Mots-clés en portugais
Imagem
Instrumentação
Ressonância magnética
Transdutores para imagem
Resumé en portugais
O objetivo deste trabalho foi melhorar a relação sinal-ruído de um sistema de imagens por ressonância magnética nuclear (TORM 005), através do desenvolvimento de bobinas de recepção mais eficientes, de circuitos de sintonia balanceados, circuitos de desacoplamento de potência para as bobinas receptoras e da melhoria do ambiente eletromagnético no qual os transdutores trabalham. Para isso, houve necessidade de se modificar a blindagem eletromagnética interna ao TORM 005 e incorporar filtros nas linhas de sinal mais críticas para o sistema. Novas bobinas para imagem foram desenvolvidas, melhorando a relação sinal-ruído e permitindo a execução de exames clínicos de diversas anatomias com qualidade superior. Sistemas de sintonia balanceados para as bobinas foram estudados e implementados, garantindo uma redução nas perdas destas bobinas devido a presença de um paciente em seu interior e permitindo uma maior reprodutibilidade da sintonia delas quando um paciente é substituído por outro. Circuitos para proteção da entrada do receptor e para desacoplamento da bobina receptora durante os pulsos de RF na transmissão foram desenvolvidos, construídos e incorporados ao TORM 005. O conjunto destas implementações garantiram ao TORM 005 uma melhora na relação sinal-ruído, na confiabilidade de operação e na reprodutibilidade dos resultados e conseqüentemente uma melhor qualidade da imagem,, o que tem permitido o uso constante do aparelho para aplicações clínicas e no desenvolvimento de novos trabalhos relacionados com imagem por RMN em ultra-baixo campo magnético.
Titre en anglais
Mri coil design and receiver system for ultra low field magnetic resonance imaging
Mots-clés en anglais
Magnetic resonance imaging
MRI coil
Nuclear magnetic resonance
Resumé en anglais
The aim of the present work was to optimize the signal to noise ratio in our NMR imaging system (TORM 005) by improving transducers' reception quality through better designed coils, balanced tuning circuit for this coils and power decoupling circuits and by reducing interference from the electromagnetic environment. For this purpose, we had to modify the internal electromagnetic shielding and incorporate line filters in the more critica1 signal paths. Also, new types of coils were developed, improving the signal to noise ratio, and allowing us to make clinical exams with superior quality for severa1 anatomies. Balanced circuits for tuning and matching of the coil were studied and built, allowing a reduction of the coil losses because patient's load. This produced a more reliable coil tuning after positioning each new patient. Circuits to avoid the receiver input overload and decoupling circuits for the isolation of receiver coils from excitation coil were designed and incorporated to the TORM 005. All these alterations of our imaging system (TORM 005) contributed to a significant improvement in the signal to noise ratio, reliability and reproducibility of the system. This permitted to operate the system routinely for clinical applications, research and development in the area of ultra low magnetic field tomography.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
EdsonVidotoM.pdf (4.44 Mbytes)