Doctoral Thesis
DOI
https://doi.org/10.11606/T.76.2010.tde-06122010-141123
Document
Author
Full name
Gustavo Henrique Frigieri Vilela
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 2010
Supervisor
Committee
Oliveira, Sergio Mascarenhas (President)
Machado, Benedito Honório
Sameshima, Koichi
Tannus, Alberto
Verzola, Roberto Mario Machado
Title in Portuguese
Desenvolvimento de um sistema minimamente invasivo para monitorar a pressão intracraniana
Keywords in Portuguese
Extensômetro
Instrumentação médica
Monitoramento minimamente invasivo
PIC
Pressão Intracraniana
Abstract in Portuguese
A pressão intracraniana (PIC) é um dos principais parâmetros neurológicos em animais e humanos. A PIC é uma função da relação entre o conteúdo da caixa craniana (parênquima cerebral, líquido cefalorraquiano e sangue) e o volume do crânio. O aumento da PIC (hipertensão intracraniana) pode acarretar graves efeitos fisiológicos ou até mesmo o óbito em pacientes que não receberem rapidamente os devidos cuidados, os quais incluem o monitoramento em tempo real da PIC. Todos os métodos de monitoramento da PIC atualmente utilizados são invasivos, ou seja, é necessário introduzir um sensor de pressão no sistema nervoso central, acarretando aos pacientes riscos de infecções e traumas decorrentes do método. Neste trabalho desenvolvemos um método minimamente invasivo de monitoramento da pressão intracraniana, que consiste na utilização de sensores de deformação do tipo strain gauge fixados sobre a calota craniana. Os sinais oriundos deste sensor foram amplificados, filtrados e enviados para um computador com software apropriado para análise e armazenamento dos dados. O trabalho aqui apresentado objetivou os testes "in vivo" do sistema, onde foi utilizada mais de uma centena de animais em diversos testes, sendo que em todos os casos os resultados foram satisfatórios, apontando a eficácia do método.
Title in English
Development of a minimally invasive system to monitor the intracranial pressure
Keywords in English
ICP
Intracranial Pressure
Medical instrumentation
Minimally invasive monitoring
Strain gauge
Abstract in English
The intracranial pressure (ICP) is one of the most important neurological parameter in animals and humans. The ICP is a function of the relation between the contents of the skull (brain parenchyma, cerebrospinal fluid and blood) and the volume of the skull. The increase in ICP (intracranial hypertension) may cause serious physiological effects and death in patients that do not receive appropriate care quickly, which includes real-time monitoring of ICP. All monitoring methods currently used in ICP are invasive, ie requiring invasion of the central nervous system by a pressure sensor, causing infections and traumas risks to patients. In this work we present a new minimally invasive method to monitor the intracranial pressure. This uses strain gauge deformation sensors, externally glued on the skull. The signal from this sensor is amplified, filtered and sent to a computer with appropriate software for analysis and data storage. "In vitro" and "in vivo" experiments let to the following results: (1) Our minimally invasive system is capable of adequately monitoring the ICP. (2) The measurements are in real and online time providing excellent signal and stability. (3) Simultaneous comparison with invasive methods not only validated our results but showed increased performance. The equipment cost effective will allow the use of our system in the Public Health System, with a important social aspect of our contribution.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2010-12-07
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.