Doctoral Thesis
DOI
https://doi.org/10.11606/T.74.2012.tde-19082014-094707
Document
Author
Full name
Max Vicente
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Pirassununga, 2012
Supervisor
Committee
Costa, Ernane José Xavier (President)
Carvalho, Aparecido Augusto de
Dirani, Ely Antonio Tadeu
Franzolin Neto, Raul
Silva, Iran José Oliveira da
Title in Portuguese
Instrumentação eletrônica baseada em sensores poliméricos para medição de amônia em ambiente zootécnico
Keywords in Portuguese
Ambiência
Bem-estar animal
Polianilina
Zootecnia de precisão
Abstract in Portuguese
O presente trabalho teve como objetivo o desenvolvimento de um dispositivo para quantificar a amônia em concentrações de interesse zootécnico. Para tal, foi produzido um sensor, composto por dois eletrodos de ouro depositado em um substrato de vidro (BK7 vidro óptico) e revestido por um filme polimérico ultrafino de polianilina. Para testar o dispositivo desenvolvido, foram realizados experimentos laboratoriais expondo o mesmo a várias concentrações de amônia que possibilitou avaliação da resposta do referido dispositivo. As medições de capacitância do sensor foram realizadas utilizando uma ponte LCR e um circuito eletrônico ZSC 31210 da ZMD International. Os resultados mostram que o sensor proposto tem reversibilidade espontânea e possui resposta linear em função da concentração de amônia nas faixas de 1 a 40 ppm. Os resultados obtidos permitem concluir que a amônia produzida em ambiente de produção animal tem a possibilidade de ser monitorada e, por conseguinte, controlada por um sistema que use um sensor polimérico capacitivo de placas paralelas.
Title in English
Electronic instrumentation based on polymeric sensors for ammonia measurement in livestock environment
Keywords in English
Ambience
Animal welfare
Livestock precision
Polyaniline
Abstract in English
This study aimed at the development and tests a device for measuring ammonia concentrations in livestock production. To accomplish this purpose, a polymeric sensor was developed by two gold electrode deposited on a glass substrate (BK7 optical glass) and coated with an ultrathin polymer film of polyaniline. In order to test the developed device, laboratory experiments were performed by exposing it at various concentrations of ammonia, which enabled to evaluate the response of the device. The capacitance of the sensor was performed using a LCR bridge and an electronic circuit ZSC31210 from ZMD International. The results show that, the proposed sensor is spontaneously reversible and has a linear response of ammonia concentration in the range from 1 to 40 ppm. Thus, the obtained results in this work indicate that, the ammonia produced in the animal production environment can be monitored and, therefore, controlled by a system using a parallel plate as a capacitor polymeric sensor.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
DO6521771COR.pdf (2.08 Mbytes)
Publishing Date
2014-08-20
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.