Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.64.2013.tde-22052013-094046
Document
Auteur
Nom complet
Taciana Figueiredo Gomes
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Piracicaba, 2013
Directeur
Jury
Zagatto, Elias Ayres Guidetti (Président)
Nogueira, Ana Rita de Araujo
Raimundo Junior, Ivo Milton
Titre en portugais
Determinação espectrofotométrica de sulfato em águas de chuva em um sistema de análises químicas em fluxo envolvendo troca-iônica
Mots-clés en portugais
Análises químicas por injeção em fluxo
Espectrofotometria UV-Vis
Resina de troca iônica
Sulfato
Resumé en portugais
Um sistema de análises químicas por injeção em fluxo foi desenvolvido e aplicado à determinação espectrofotométrica de baixas concentrações de sulfato em águas de chuva. O método explora a reação de deslocamento entre o ânion sulfato e o complexo dimetilsulfonazo (III) de bário, Ba-DMSA, com consequente formação do sal pouco solúvel BaSO4. A solução com o complexo Ba-DMSA apresenta alta absortividade molar a 656 nm, e uma diminuição pronunciada na absorbância ocorre na presença de sulfato. Desta forma, o sinal analítico é registrado como pico invertido proporcional à concentração de sulfato na amostra. A presença de altas concentrações de dimetilsulfóxido, DMSO, no reagente favorece a intensidade do sinal analítico, por coordenar as moléculas de água, melhorando a sensibilidade do método. A utilização do tensoativo dodecilsulfato de sódio, SDS, determina também uma melhoria considerável na sensibilidade analítica, pois sua presença no ambiente reacional altera as condições de supersaturação. A sensibilidade e seletividade do procedimento são melhoradas incluindo-se uma etapa de troca iônica em linha, com resina aniônica Dowex 1-X8, para concentração do sulfato e separação dos íons interferentes em potencial. O sistema proposto é robusto e fornece resultados precisos (d.p.r. < 0,01) com frequência analítica de 30 determinações por hora. O consumo de DMSA é de 0,17 'mü'mol por determinação. A curva analítica se apresenta linear até 2,0 mg L-1 SO42- e os limites de detecção e de quantificação foram estimados em 0,01 mg L-1 SO42- e 0,04 mg L-1 SO42-.
Titre en anglais
Spectrophotometric determination of sulphate in rain waters using a flow injection system with ion-exchange
Mots-clés en anglais
Flow injection analysis
Ion exchange resin
Sulfate
UV-Vis spectrophotometry
Resumé en anglais
A flow injection system was developed and applied to the spectrophotometric determination of low concentrations of sulphate in rain waters. The method exploits the displacement reaction involving the sulphate anion and the barium dimethylsulfonazo (III) complex, Ba-DMSA, with consequent formation of the slightly soluble BaSO4 salt. As the Ba-DMSA solution is characterized by a high molar absorptivity at 656 nm, a pronounced absorbance lessening is observed in the presence of sulphate. The analytical signal is then recorded as an inverted peak proportional to the sulphate content in the sample. Presence of dimethylsulphoxide, DMSO, at high concentrations in the reaction medium leads to an enhancement of the analytical signals, coordinating water molecules, thus improving the sensitivity of the method. Addition of the surfactant sodium dodecylsulphate, SDS, results also in a pronounced improvement in sensitivity, as its presence in the reaction medium alters the supersaturation conditions. The sensitivity and selectivity of the method were improved by including an in-line ion exchange mini-column with the anionic resin Dowex 1-X8 into the manifold, for sulphate concentration and separation from potential interfering ions. The proposed system is robust and provides accurate results (R.S.D. <0.01) at an analytical frequency of 30 determinations per hour. About 0.17 'mü'mol DMSA are consumed per determination. The analytical curve is linear up to 2.0 mg L-1 SO42- and the detection and quantification limits were estimated as 0.01 mg L-1 SO42- and 0.04 mg L-1 SO42-.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2013-05-29
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.