Disertación de Maestría
DOI
https://doi.org/10.11606/D.64.2014.tde-03112014-150442
Documento
Autor
Nombre completo
Andressa Adame
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Piracicaba, 2014
Director
Tribunal
Rocha, Fabio Rodrigo Piovezani (Presidente)
Masini, Jorge Cesar
Fatibello Filho, Orlando
Título en portugués
Sistema de análises em fluxo empregando multicomutação para avaliação de toxicidade aguda com Vibrio fischeri
Palabras clave en portugués
Análises em fluxo
Diluições em Linha
Ecotoxicidade
Multi-impulsão
Multicomutação
Poluentes emergentes
Resumen en portugués
Um sistema de análises em fluxo com multicomutação foi desenvolvido para a mecanização do bioensaio de toxicidade aguda com a bactéria Vibrio fischeri, o qual é baseado na diminuição da luminescência bacteriana causada por espécies tóxicas.Microbombas solenoide foram utilizadas para melhorar as condições de mistura, permitindo também que diluições fossem realizadas em linha com a alteração do volume de amostra e a sobreposição parcial entre a zona da amostra e a suspensão bacteriana. Uma cela de fluxo em espiral, posicionada na janela de emissão de um espectrofluorímetro, foi empregada para as medidas de luminescência e um banho de água construído em laboratório, baseado no efeito Peltier, foi utilizado para manter a temperatura do bioensaio em (13,0±0,1) °C. Um pequeno volume da suspensão bacteriana (120 ?L) foi selecionado visando minimizar o consumo e, consequentemente, os custos do ensaio. O tempo de contato entre a suspensão bacteriana e a amostra foi reduzido para 5 min para aumentar a frequência de amostragem e evitar a atenuação da luminescência devido ao tempo de vida curto das bactérias. Os valores de CE50 para os controles positivos foram estimados em 2,5 ± 0,7, 2,0 ± 0,5 e 10 ± 4 mg L-1 para o Zn(II), Cu(II), e Cr(VI), respectivamente, sendo concordantes com os valores da literatura. Os valores de CE50 para alguns poluentes emergentes (parabenos, cafeína, acetaminofeno, diclofenaco e ácido salicílico) concordaram com os obtidos utilizando o kit comercial BioTox(TM), realizado em micro placa, a nível de confiança de 95 %. Os coeficientes de variação foram estimados em 2,4 e 2 % (n = 10), na ausência e na presença de 0,6 mg L-1 de Zn(II), respectivamente. O sistema de análises em fluxo proposto é, então, uma alternativa simples, rápida, robusta e precisa para a determinação da toxicidade aguda utilizando pequenos volumes de amostra e suspensão bacteriana. Além disso, o sistema apresenta vantagens em comparação com os bioensaios em batelada e em fluxo, tais como ajuste de salinidade e diluições das amostras em linha e a determinação da cinética de toxicidade das espécies químicas
Título en inglés
A multicommuted flow-based system for acute toxicity evaluation with Vibrio fischeri
Palabras clave en inglés
Ecotoxicity
Emerging pollutants
Flow analysis
In-line dilutions
Multi-pumping
Multicommutation
Resumen en inglés
A multicommuted flow system was developed for automation of the acute toxicity bioassay using Vibrio fischeri bacteria, which exploits the decrease of the bacterial luminescence caused by toxic species. Solenoid micro-pumps were employed to improve mixing conditions and to perform on-line dilutions of the tested compounds by changing the sample volume and exploiting the partial overlap between sample zone and the bacterial suspension. A spiral flow cell, placed at the emission window of a spectrofluorimeter, was employed for signal measurement and a lab-made water bath based on the Peltier effect was used for maintain the temperature of the bioassay at (13.0±0.1) oC. A low volume of the bacterial suspension (120 ?L) was selected in order to minimize its consumption and, consequently, the costs of the assay. The contact time between bacterial suspension and sample was reduced to 5 min to increase the sampling rate and to avoid luminescence fading due to the bacteria short lifetime. EC50 values for positive controls were estimated as 2.5 ± 0.7, 2.0 ± 0.5 and 10 ± 4 for Zn(II), Cu(II), and Cr(VI), respectively, being in agreement with literature values. The EC50 values for some emerging pollutants (parabens, caffeine, acetaminophen, diclofenac and salicylic acid) agreed with those obtained from commercial BioTox(TM) kit performed in micro plate at the 95% confidence level. The coefficients of variation were estimated as 2.4 and 2 % (n=10), in the absence and presence of 0.6 mg L-1 Zn(II), respectively. The proposed flow system is then a simple, fast, robust and accurate alternative for acute toxicity determination, using low sample and bacterial suspension volumes. Furthermore, the system presents advantages in comparison to batch and previous flow-based bioassays, such as the achievement of in-line salinity adjustment and sample dilutions, and the determination of toxicity kinetic for every assayed chemical species
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
AndressaAdame_Revisada.pdf (1,000.68 Kbytes)
Fecha de Publicación
2014-12-01
 
ADVERTENCIA: El material descrito abajo se refiere a los trabajos derivados de esta tesis o disertación. El contenido de estos documentos es responsabilidad del autor de la tesis o disertación.
  • ADAME, A., e ROCHA, F. R. P. A multi-pumping flow system for acute toxicity bioassay using the Vibrio fischeri bacteria [doi:10.1039/c4ay00789a]. Analytical Methods [online], 2014, vol. 6, p. 7367-7373.
  • ADAME, A., e ROCHA, F. R. P. A multicommuted flow-based system for acute toxicity bioassay using Vibrio fischeri bacteria. In 18th International Conference on flow Analysis, Porto-Portugal, 2013. 18th International Conference on flow Analysis., 2013. Resumo.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.