Doctoral Thesis
DOI
https://doi.org/10.11606/T.59.2018.tde-31052018-223216
Document
Author
Full name
Vinicius Ricardo Acquaro Junior
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2018
Supervisor
Committee
Nassur, Maria Eugenia Queiroz (President)
Augusto, Fabio
Gaitani, Cristiane Masetto de
Martinis, Bruno Spinosa de
Rocha, Eduardo Carasek da
Title in Portuguese
Desenvolvimento de diferentes métodos LC-MS/MS para a determinação de fármacos e endocanabinóides em amostras de plasma
Keywords in Portuguese
Bio-SPME-Nano-ESI-MS/MS; Column switching; Colunas superficialmente porosas; Endocanabinóides; Esquizofrenia; Parâmetros cromatográficos; SPME-UHPLC-M
Abstract in Portuguese
Esta tese foi dividida em três capítulos. O capítulo I descreve o desenvolvimento do método Column switching UHPLC-MS/MS para a determinação simultaneamente de fármacos psicotrópicos em amostras de plasma de pacientes esquizofrênicos. A politerapia é uma prática comum no tratamento da esquizofrenia. Portanto, a monitorização terapêutica destes fármacos tem sido realizada para o ajuste das doses e individualização da terapia farmacológica. O método Column switching UHPLC-MS/MS apresentou linearidade na faixa de concentração de 0,025 a 1,25 ng mL-1 com R2 acima de 0,9950 e a falta de teste de ajuste (p > 0,05); precisão com coeficientes de variação inferiores a 12% e exatidão com erro padrão relativo inferior a 14%. Este método foi aplicado com sucesso para determinação de fármacos em amostras de plasma de pacientes esquizofrênicos para fins de monitorização terapêutica. No capítulo II, o desempenho cromatográfico de colunas C18 superficialmente e totalmente porosas com diferentes tamanhos de partícula foi avaliado para a análise de fármacos psicotrópicos por LC-MS/MS e LC-DAD. Com o sistema LC-MS/MS foram avaliados os seguintes parâmetros cromatográficos: altura do prato reduzido vs velocidade linear reduzida, impedância vs velocidade linear reduzida, tempo da corrida cromatográfica vs vazão, pressão vs vazão, resolução, capacidade de pico, assimetria e fator de retenção. Já com o sistema LC-DAD foram avaliados a hidrofobicidade, atividade silanol e impurezas metálicas também foram avaliadas. As colunas com superfície carregada apresentaram maior eficiência cromatográfica para os fármacos em sua forma ionizada. Já as colunas com partículas menores que 2 µm (Cortecs 1,6 µm, Acquity 1,7 µm, e Kinetex 1,7 µm) apresentaram maior eficiência cromatográfica para os fármacos na forma parcialmente ionizada. Os modelos matemáticos gerados foram capazes de prever a pressão e o tempo da corrida cromatográfica em diferentes vazões para todas as colunas. Considerando a eficiência, impedância, resolução, capacidade de pico, fator de retenção e hidrofobicidade, as colunas Cortecs 1,6 µm e Acquity 1,7 µm apresentaram melhor desempenho durante a análise dos fármacos em amostra de plasma. O capítulo III descreve o desenvolvimento e validação dos métodos SPME-UHPLC-MS/MS e Bio-SPME-Nano-ESI-MS/MS para a determinação dos endocanabinóides (AEA e 2-AG) em amostras biológicas. Para a otimização do processo SPME foram avaliadas as fases SPME (C18, C30 e HLB) e os solventes para dessorção (metanol, acetonitrila e isopropanol). Os aditivos modificadores de matriz, como cloridrato de guanidina, ácido trifluoroacético e acetonitrila foram avaliados por planejamento experimental. Os métodos SPME-UHPLC-MS/MS e Bio-SPME-Nano-ESI-MS/MS, com a fase HLB biocompatível, apresentaram para ambos endocanabinóides valores de LOQs de 1 ng mL-1 e 50 ng mL-1, respectivamente. O método Bio-SPME-Nano-ESI-MS/MS permitiu o direto acoplamento da fibra SPME ao espectrômetro de massas via dessorção/ionização nanoeletrospray que resultou em rápida determinação quantitativa dos endocanabinóides em amostras biológicas.
Title in English
Development of different LC-MS/MS methods for the determination of drugs and endocannabinoids in plasma samples
Keywords in English
Bio-SPME-Nano-ESI-MS/MS; Chromatographic parameters; Column switching; Endocannabinoids; Schizophrenia; SPME-UHPLC-MS/MS; Superficially porous columns
Abstract in English
This thesis is divided into three chapters. Chapter I describes the development of a column switching UHPLCMS/MS method to determine psychotropic drugs in schizophrenic patients plasma samples simultaneously. Polytherapy is a common practice in schizophrenia treatment. Therefore, therapeutic drug monitoring has been applied to adjust doses and to customize pharmacological therapy. The column switching UHPLCMS/MS method developed here is linear at concentrations ranging from 0.025 to 1.25 ng mL-1 with R2 above 0.9950 and presents lack of fit test (p > 0.05), precision with coefficients of variation lower than 12%, and accuracy with relative standard error lower than 14%. This method was successfully applied to determine drugs in schizophrenic patients plasma samples for therapeutic drug monitoring. In chapter II, the chromatographic performance of C18 superficially porous columns and of C18 fully porous columns with different particle sizes were evaluated for analysis of psychotropic drugs by LC-MS/MS and LC-DAD. Within the LC-MS/MS system, the following chromatographic parameters were assessed: reduced plate height vs reduced linear velocity, impedance vs reduced linear velocity, chromatographic run time vs flow rate, backpressure vs flow rate, resolution, peak capacity, asymmetry, and retention factor. Within the LC-DAD system, hydrophobicity, silanol activity, and metal impurities were also examined. Columns with charged surface displayed improved chromatographic efficiency for drugs in the ionized form. Columns with particles smaller than 2 µm (Cortecs 1.6 µm, Acquity 1.7 µm, and Kinetex 1.7 µm) presented higher chromatographic efficiency for the drugs, which were in their partially ionized form. The generated mathematical models were able to predict the backpressure and the chromatographic run time at different flow rates for all the columns. Considering efficiency, impedance, resolution, peak capacity, retention factor, and hydrophobicity, columns Cortecs 1.6 µm and Acquity 1.7 µm provided the best performance during analysis of drugs in plasma samples. Chapter III describes the development and validation of the SPME-UHPLC-MS/MS and the Bio-SPME-Nano-ESI-MS/MS methods for determination of endocannabinoids (AEA and 2-AG) in biological samples. To optimize the SPME process, SPME coatings (C18, C30, and HLB) and solvents for desorption (methanol, acetonitrile, and isopropanol) were evaluated. Matrix modifier additives, such as guanidine hydrochloride, trifluoroacetic acid, and acetonitrile, were assessed by experimental design. The SPME-UHPC-MS/MS and the Bio-SPME-Nano-ESI-MS/MS methods with HLB biocompatible coating provided LOQ values of 1 ng mL-1 and 50 ng mL-1, respectively, for both endocannabinoids. The Bio-SPME-Nano-ESI-MS/MS method allowed direct coupling of SPME fibers to the mass spectrometer by desorption/ionization nanoelectrospray, which resulted in rapid quantitative determinations of endocannabinoids in biological samples.
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2018-07-24