Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.59.1999.tde-23022024-111921
Document
Auteur
Nom complet
Felipe Chen Abrego
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 1999
Directeur
Jury
Baffa Filho, Oswaldo (Président)
Martin Neto, Ladislau
Rossi, Alexandre Malta
Titre en portugais
Estudo de otimização do registro de espectros de ressonância paramagnética eletrônica de dosímetros de alanina em baixas doses
Mots-clés en portugais
Dosimetria
Espectrômetro de ressonância paramagnética eletrônica
Espectros de EPR
Física médica
Resumé en portugais
Realizou-se um estudo da otimização dos parâmetros de operação de um espectrômetro de ressonância paramagnética eletrônica para o registro de espectros de EPR de dosímetros de DL e L-alanina irradiados com radiação gama do 60Co, visando estabelecer-se um procedimento de dosimetria Alanina/EPR capaz de detectar na faixa de doses terapêuticas (~ 1 Gy). Para lograr o anteriormente mencionado, fabricaram-se dosímetros com 80% em peso de alanina mais 20% em peso de parafina e com dimensões nominais de 4,7 mm de diâmetro e 12 mm de comprimento. Estudou-se a variação da amplitude do espectro de EPR de pastilhas irradiadas com a potência de microondas e com a amplitude de modulação do campo magnético. Posteriormente, registraram-se espectros de EPR do primeiro e segundo harmônico de pastilhas irradiadas e não irradiadas com uma amplitude de modulação de 10 G e com um campo de varredura de 40 G, encontrando-se que para este valor de modulação, a amplitude do segundo harmônico não é desprezível em comparação ao primeiro harmônico e representa uma contribuição em amplitude ao redor de 38%. Na faixa de 1-20 Gy, a L-alanina mostrou ser 20% mais sensível que a DL-alanina. Na faixa de 0,05-1Gy a sensibilidade de ambas é semelhante apresentado uma diferença de somente 3%. A amplitude do sinal de EPR para as pastilhas não irradiadas representou uma dose de 0,5 Gy e este valor estabelece o limite inferior de detecção do sistema.
Titre en anglais
Study to optimize the recording of electronic paramagnetic resonance spectra of alanine dosimeters at low doses
Mots-clés en anglais
Dosimetry
Electronic paramagnetic resonance spectrometer
EPR spectra
Medical physics
Resumé en anglais
An optimization of the operation parameters for an electron paramagnetic resonance (EPR) spectrometer was made for the acquisition of EPR signals from DL and L-alanine dosimeters irradiated with 60Co gamma radiation, with the aim of establish an Alanine/EPR dosimetry procedure capable to detect doses in the therapeutic range (~ 1 Gy). To attain this aim, dosimeters with 80% weight of alanine and 20% weight of paraffin and nominal dimensions: 4,7 mm diameter and 12 mm length were made. A study of the variation of the EPR signal amplitude from irradiated alanine pellets with the microwave power and with the magnetic field modulation amplitude was made. Later, EPR signals of first and second harmonic from irradiated and no-irradiated pellets were recorded with 10 G (1 mT) modulation amplitude and 40 G (4 mT) of scan width. With this overmodulation value was found that, the amplitude of the second harmonic is not small in comparison with the amplitude of the first harmonic and represent approximately 38% of the amplitude of the first harmonic. ln the dose range of 1-20 Gy, the L-alanine was 20% more sensitive than the DL-alanine. ln the dose range 0,05-1 Gy, the sensitive of both were similar showing a 3% difference only. The EPR signal amplitude from non-irradiated pellets represented a dose of 0,5 Gy, and this value fix the lower limit of detection for the Alanine/EPR dosimetric system.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
001063726.pdf (9.24 Mbytes)
Date de Publication
2024-02-23
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.