• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.58.2012.tde-23052012-150900
Documento
Autor
Nombre completo
Rodrigo Teixeira Macri
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Ribeirão Preto, 2012
Director
Tribunal
Matsumoto, Mirian Aiko Nakane (Presidente)
Borsatto, Maria Cristina
Romano, Fábio Lourenço
Sakima, Mauricio Tatsuei
Sinhoreti, Mário Alexandre Coelho
Título en portugués
Laser de CO2 como auxiliar na remoção de bráquetes estéticos cerâmicos
Palabras clave en portugués
braquetes ortodônticos cerâmicos
Descolagem
lasers de CO2
Resumen en portugués
Este estudo avaliou a eficiência do uso do laser de CO2 com comprimento de onda de 10,6 m como auxiliar na remoção de bráquetes estéticos cerâmicos. Para este estudo foram utilizados 105 pré-molares inferiores humanos recentemente extraídos, que receberam irradiações com laser de CO2 antes da descolagem dos bráquetes. Na primeira fase deste estudo, o objetivo foi mensurar e comparar a temperatura no compósito de colagem do bráquete e na câmara pulpar após aplicação do laser de CO2 , para observar a segurança para o esmalte, bem como se a temperatura era suficiente para fragilizar o compósito, facilitando a remoção do bráquete. Nesta etapa foram utilizados 30 pré-molares inferiores recentemente extraídos, sendo que cada um recebeu 4 irradiações com diferentes protocolos de maneira aleatória. Foram testados 12 diferentes protocolos variando o tempo de irradiação (3 e 5s), a duração do pulso (0,01 e 0,03s) e a potência (5, 8 e 10W). Cada protocolo foi utilizado por 10 vezes (n=10), totalizando 120 irradiações. Em todas as irradiações os corpos-de-prova foram imersos em banho térmico com água a 37º C. Foram encontrados 4 protocolos cuja variação da temperatura na polpa foi menor que 5,5º C, portanto compatível com a saúde da polpa. Na segunda etapa do estudo, o objetivo foi avaliar a resistência ao cisalhamento após a aplicação do laser utilizando os protocolos cuja temperatura na polpa não fosse superior a 5,5º C. Setenta e cinco pré-molares inferiores humanos foram divididos em 5 grupos, dos quais um não foi submetido à irradiação, sendo considerado Controle (C). Nos 4 grupos restantes, irradiações foram realizadas como descrito a seguir: Grupo I (potência 5W; tempo de pulso 0,01s; tempo de aplicação 3s), Grupo II (5W; 0,03s e 3s), Grupo III (8W; 0,01s e 3s) e Grupo IV (10W; 0,01s e 3s). Cada grupo foi formado por 15 corpos-de-prova (n=15). Após a descolagem do bráquete foi avaliado o Índice de Remanescente Adesivo (IRA). Após as mensurações e aplicação da Análise de Variância (ANOVA) e do teste de Tukey foi constatado que o único grupo que apresentou diminuição na força de remoção foi o IV (7,33 MPa) em relação ao grupo C (10,81 MPa) sendo assim considerado seguro e eficaz na diminuição da resistência do compósito.
Título en inglés
CO2 laser-assisted debonding for esthetic ceramic brackets
Palabras clave en inglés
ceramic orthodontic brackets
CO2 laser
debonding
Resumen en inglés
This study evaluates the efficacy CO2 10,6 m laser-assisted debonding esthetic ceramic brackets. For this purpose it was used 105 recently extracted lower human premolars that were irradiated with CO2 laser before debonding. In the first phase, the objectives were to measure and compare the temperature in composite for bracket bonding and at the pulp chamber after CO2 laser irradiation, to observe the safety for enamel, as well the temperature was enough to softening the composite, making debonding bracket easier. In this part it was used 30 lower premolar recently extracted, and each one received 4 irradiations with different protocols, randomly choose. It was tested 12 different protocols varying time of irradiation (3 and 5s), pulse duration (0,01s e 0,03s) and power (5, 8 and 10W). Each protocol was repeated for 10 times (n=10), resulting 120 irradiations. In every irradiation the teeth tested were immerse in warm water bath at 37o C. It was found 4 protocols with pulp temperature lower than 5,5º C, so this is compatible with pulp health. In the second phase of this study, the objective was to evaluate the shear bond strength after laser irradiation, using the protocols with pulp chamber temperatures lower than 5,5º C. Seventy-five lower human premolars was shared in 5 groups. One group did not receive laser irradiation and considered Control (C) and the other groups the irradiation was made as it follows: Group I (power density 5W; pulse duration 0,01s; irradiation time 3s), Group II (5W; 0,03s, 3s), Group III (8W; 0,01s; 3s) and Group IV (10W; 0,01s; 3s). After results ANOVA and Tukey tests were made. Group IV presented lowering in SBS force (7,33 MPa) when compared with C (10,81MPa). This protocol was founded as a safe and efficient in softening composite layer.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2012-09-25
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.