Titre en portugais
Avaliação da resistência ao arrancamento e do torque de remoção de um novo modelo de implante de zircônia
Mots-clés en portugais
Ciclagem termomecânica
Ensaio de arrancamento
Estabilidade primária
Implantação dentária
Titânio
Torque de remoção
Zircônia dopada com ítria
Resumé en portugais
O objetivo deste estudo foi avaliar comparativamente a estabilidade primária e a longevidade de implantes de zircônia (Y-TZP) e titânio (Grau IV) após ciclagem termomecânica (CTM). Foram obtidos 40 implantes de cada material que foram instalados em osso artificial e separados em grupos (n=10) segundo o material, a realização de ciclagem termomecânica e o ensaio realizado: Titânio - G1 (ensaio de torque de remoção); G2 (ensaio de arrancamento); G3 (CTM + ensaio de torque de remoção); G4 (CFTM + ensaio de arrancamento); e Zircônia - G5 (ensaio de torque de remoção); G6 (ensaio de arrancamento); G7 (CTM + ensaio de torque de remoção); G8 (CTM + ensaio de arrancamento). O equipamento de ciclagem mecânica foi configurado com carga de 133 N, frequência de 120 ciclos por minuto (2 Hz), totalizando 2.000.000 ciclos em cada espécime. A ciclagem térmica foi realizada com temperatura entre 5ºC, 37ºC e 55ºC. Após os ensaios mecânicos, os resultados foram submetidos à análise estatística (2-way ANOVA, teste de Bonferroni, p<0,05) e verificou-se que houve diferença estatisticamente significante (p<0,05) entre os implantes de titânio e zircônia, independente do tratamento utilizado, tanto para o torque de remoção quanto para a foça máxima de arrancamento. Os implantes de titânio mostraram maiores valores de torque de remoção e força de arrancamento comparado aos implantes de zircônia (p<0,05). Verificou-se também que a ciclagem termomecânica foi significante apenas para o ensaio de arrancamento quando utilizados implantes de titânio. Conclui-se que a estabilidade primária de implantes de titânio foi maior do que implantes de zircônia e que a ciclagem termomecânica foi um fator significativo apenas para a longevidade dos implantes de titânio.
Titre en anglais
Analysis of pullout strength and removal torque of a new zirconia dental implants
Mots-clés en anglais
dental implantation
primary stability
pull-out test
removal torque
thermomechanical cyclic
titanium
yttrium-doped zirconia
Resumé en anglais
The aim of this study was to assess comparatively the primary stability and longevity of zirconia (Y-TZP) and titanium (Grade IV) implants after thermomechanical cyclic (CTF). Forty implants to each material were obtained, installed in artificial bone and separated into groups (n=10) according to the material, the conducting of thermomechanical cyclic and the test submitted: Titanium - G1 (removal torque); G2 (pullout test); G3 (CTF + removal torque); G4 (TCF + pullout test); Zirconia - G5 (removal torque); G6 (pullout test); G7 (CTF + removal torque); G8 (CTF + pullout test). The mechanical cyclic testing machine was programmed to apply a load of 133 N, frequency of 120 cycles per minute (2 Hz) and a total of 2,000,000 cycles in each specimen. The thermocycling was set with a temperature between 5°C, 37ºC and 55ºC. After the mechanical tests, the data were subjected to statistical analysis (2-way ANOVA, Bonferroni test, p <0,05) and the results showed statistically significant difference (p <0.05) between zirconia and titanium implants, regardless of treatment used, for both removal torque and pullout tests. Titanium implants showed higher removal torque and pullout strength, which was statistically significant (p <0,05) compared with zirconia implants. It was also found that the thermomechanical cyclic was only significant for pullout test when used titanium implants. It was concluded that primary stability of titanium implants is higher than zirconia implants and the thermomechanical cyclic was a significant factor only for the longevity of titanium implants.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2017-08-17
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.