Master's Dissertation
DOI
https://doi.org/10.11606/D.54.1984.tde-19052014-154254
Document
Author
Full name
Edson dos Santos Moreira
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 1984
Supervisor
Committee
Slaets, Jan Frans Willem (President)
Craievich, Aldo Felix
Linhares, Odelar Leite
Title in Portuguese
Desenvolvimento de um analisador de altura de pulsos
Keywords in Portuguese
Instrumentação eletrônica
Microprocessador
Processamento digital de dados
Visualização de dados
Abstract in Portuguese
Neste trabalho descrevemos o desenvolvimento de um analisador de altura de pulsos. Este aparelho é essencial no estudo de sinais oriundos de sensores que detectam fenômenos físicos e codificam as informações na amplitude dos pulsos que fornecem na saída. O sistema compõe-se de um módulo de entrada de sinais conectado a um módulo de controle baseado no microprocessador 8085ª capaz de memorizar pulsos com até 1 uS de largura em 256 canais com resolução melhor que 20mV. Um módulo de comunicação dotado de interface serial é usado para transferência de dados para outros dispositivos através do protocolo RS232c. O módulo de operação e monitoração munido de teclado hexadecimal e saída analógica possibilita a visualização das curvas coletadas num monitor XY. A arquitetura do aparelho e os programas desenvolvidos para este sistema de baixo custo foram otimizados de forma a produzir um tempo morto típico de aproximadamente 100 uS. Como aplicação ele foi utilizado para levantamento de curvas no Laboratório de espelhamento de raios-x a baixo ângulo deste Departamento. O desempenho do aparelho foi testado através de comparações entre seus dados e os obtidos através de um similar, um PHA Northern modelo NS633, e conclui-se pela sua eficiência
Title in English
The development of pulse height analyzer
Keywords in English
Data visuallization
Digital data processing
Electronic instrumentation
Microprocessor
Abstract in English
This work describes the development of a Pulse height analyzer. This equipment is essential to analyze data coming from detectors producing information codified in pulse amplitudes. The system developed consist of a Signal input module connected to a controller module based on a 8085A microprocessor capable to memorize pulses up to 1 us in 256 channels with a resolution better than 20mV. A Communication module with a serial interface is used for data transfer to a host computer using RS232c protocol. The monitoring and operation module consist of a hexadecimal Keyboard, a 6 digit 7-segment display and a XY monitor. The hardware and the software designed for this low cost system were optimized to obtain a typical dead time of approximately 100 uS. As application, this device was used to acquire curves at the Small Angle x-ray scattering laboratory in this Department. The apparatus performance was tested by comparing its data with a Northern Pulge height analyzer model NS633 output, with favorable results
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-05-20
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.