• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Tesis Doctoral
DOI
https://doi.org/10.11606/T.46.2013.tde-20032014-090601
Documento
Autor
Nombre completo
Eric Tavares da Costa
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2013
Director
Tribunal
Gutz, Ivano Gebhardt Rolf (Presidente)
Jesus, Dosil Pereira de
Masini, Jorge Cesar
Moraes, Maria de Lourdes Leite de
Oliveira, Marcone Augusto Leal de
Título en portugués
Desenvolvimento de um sistema de análise de compostos voláteis ionizáveis por eletroforese capilar combinando técnicas de microfabricação
Palabras clave en portugués
Agentes neurotóxicos
Eletroforese capilar
Laser de dióxido de carbono
Microfluídica
Robô
Resumen en portugués
O desenvolvimento de um equipamento de eletroforese capilar com detecção condutométrica sem contato e hifenizado a um sistema de amostragem de gases é apresentado. Ele foi construído a partir de chapas de poli(metacrilato de metila) por ablação a laser de CO2, mesclando métodos convencionais com os de microfabricação. Este protótipo traz como inovações: (1) a refrigeração em estado sólido da coluna capilar, (2) a inclusão de eletrólise separada em microdispositivos e (3) a montagem de uma válvula embutida no microchip, baseada em servomotor. O equipamento é portátil (31 × 29 × 25 cm) e projetado para ser utilizado em bancada ou embarcado num robô para análises remotas de agentes neurotóxicos e outros gases ionizáveis em contato com a fase aquosa. O sistema de termostatização permitiu o controle da temperatura interna do capilar de sílica fundida entre 12 e 50 °C utilizando uma elemento Peltier a a partir da temperatura ambiente de 21 °C. A amostragem de gases é feita utilizando-se uma membrana tubular porosa acoplada ao sistema microfluídico. Ela possui 10 cm de comprimento com 300 micrometros de diâmetro externo e 3,1 microlitros de volume interno. Como prova de conceito, foram feitas amostragens de ar contendo ácidos fórmico, acético e propiônico, separação por eletroforese capilar dos respectivos ânions e detecção com relação sinal-ruído de 414, 150 e 115, respectivamente. A amostragem foi realizada por 30 s e a injeção eletrocinética, por 2 s a 1 kV. A corrida eletroforética foi concluída em 4 min.
Título en inglés
Development of an analysis system for ionizable gases using capillary electrophoresis combining conventional and microfabrication techniques
Palabras clave en inglés
Capillary electrophoresis
Carbon dioxide laser
Microfluidics
Neurotoxic agents
Robot
Resumen en inglés
The development of a capillary electrophoresis equipment with contactless conductivity detection coupled to a gas sampling system is described. It was constructed from poly(methyl methacrylate) by carbon dioxide laser ablation, using conventional methods and microfabrication. The innovations are: (1) solid-state cooling system of the silica capillary, (2) the inclusion of separate electrolysis on a microdevices, and (3) a valve assembly embedded in the microchip based servomotor. It is portable (31 × 29 × 25 cm) and designed to be used on a workbench or embedded in a robot for remote analysis of nerve agents and other ionized gases. The thermostating system allowed the temperature control of the fused silica capillary between 12 and 50 °C using a Peltier chip from ambient temperature of 21°C. Gas sampling is performed using a tubular porous membrane coupled to the microfluidic system. It is 10 cm long with outer diameter of 300 micrometers and 3.1 microliters of internal volume. Air containing formic, acetic, and propionic acids were sampled, and the corresponding anions were separated and detected in the equipment as a proof of concept. The signal-to-noise ratio were, respectively, 414, 150, and 115, after sampling by 30 s and electrokinetic injection by 2 s at 1 kV. The electrophoretic run was accomplished in 4 min.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-04-08
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.