Tesis Doctoral
DOI
https://doi.org/10.11606/T.46.2001.tde-17102002-122859
Documento
Autor
Nombre completo
José Alberto Fracassi da Silva
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2001
Director
Tribunal
Lago, Claudimir Lucio do (Presidente)
Carrilho, Emanuel
Gutz, Ivano Gebhardt Rolf
Pasquini, Celio
Tavares, Marina Franco Maggi
Título en portugués
Detecção condutométrica sem contato (oscilométrica) para eletroforese capilar de zona e cromatografia micelar eletrocinética
Palabras clave en portugués
detecção eletroquímica
eletroforese capilar
instrumentação
Resumen en portugués
Este trabalho descreve a construção e avaliação de um detector condutométrico sem contato (oscilométrico) para sua aplicação em eletroforese capilar de zona e cromatografia micelar eletrocinética (MEKC). A construção do detector contou com a avaliação de diversos materiais e métodos para a confecção dos eletrodos, tão bem como o aperfeiçoamento do seu circuito eletrônico. O seu comportamento e desempenho foram verificados através do estudo dos diversos parâmetros que influenciam sua resposta, como freqüência e amplitude do sinal aplicado, temperatura e condutividade do meio. Além disso, a simulação do circuito equivalente da cela de detecção auxiliou no entendimento das propriedades do detector frente a alterações na condutividade do meio, na freqüência de operação e nas dimensões da cela. A otimização dos parâmetros operacionais foi racionalizada pela formulação de equações analíticas que descrevem o fator de resposta do detector a partir de parâmetros obtidos experimentalmente. Para o desenvolvimento do sistema de detecção, dois equipamentos completos de eletroforese capilar foram construídos. Sistemas de injeção de amostra por pressão, por gravidade, e eletrocinética foram desenvolvidos. Um dos equipamentos permite que a injeção da amostra seja feita do lado aterrado da fonte de alta tensão. Conseqüentemente, o detector é posicionado próximo do ponto de aplicação da alta tensão. Este é, de que se tem notícia, o primeiro sistema de detecção eletroquímico para eletroforese capilar com essa característica. A termostatização dos capilares foi efetuada por convecção forçada de ar. De maneira a aumentar o grau de automação e facilitar o seu manuseio, os dois equipamentos foram interfaceados a microcomputador. Os equipamentos de eletroforese com detecção oscilométrica construídos permitiram a análise de diversas substâncias em matrizes variadas, como por exemplo íons orgânicos e inorgânicos em água de chuva e material particulado, ácidos graxos de cadeia linear em gordura de coco, formaldeído na fase gasosa da atmosfera, e, pela primeira vez, espécies neutras, como álcoois alifáticos, separadas por MEKC.
Título en inglés
Contactless conductivity detection for capillary zone electrophoresis and micellar electrokinetic chromatography
Palabras clave en inglés
capillary electrophoresis
electrochemical detection
instrumentation
Resumen en inglés
This work describes the construction and evaluation of a contactless conductometric (or oscillometric) detector, for its application in capillary zone electrophoresis and micellar electrokinetic chromatography (MEKC). The materials and methods used for the construction of the electrodes, as well as the improvements on the electronic circuit, were evaluated. The behavior and performance of the detector were studied by changing the parameters that affect its response, e. g., the frequency and amplitude of the input signal, temperature and solution conductivity. The simulation of the cell equivalent electric circuit helped to understand the detector properties related to the variations in the solution conductivity, frequency, and cell dimensions. The optimization of the parameters was conducted by the formulation of analytical equations that describe the response factor using experimental data. For the detection development, two complete capillary electrophoresis equipments were constructed. Pressure, gravity, and electrokinetic sample injection systems were developed. One of the equipments allows the sample injection on the grounded side of the capillary. Consequently, the detector is placed near the high voltage application point. This is the first electrochemical detector described for capillary electrophoresis with this characteristic. The capillary thermostating was made by passing a forced air stream. Both equipments were interfaced to microcomputers. The capillary electrophoresis equipments with oscillometric detection were applied to the analysis of many types of compounds in different matrices, such as organic and inorganic ions in rainwater and particulate material, fatty acids in cocoa oil, formaldehyde in the atmosphere and, for the first time, neutral species, like aliphatic alcohols, separated by MEKC.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fracassi.pdf (4.85 Mbytes)
Fecha de Publicación
2003-08-14
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.