Título en portugués
Correlação experimental de ângulo sólido para correlação angular gama-gama
Palabras clave en portugués
Correção de ângulo sólido
Correlação angular gama-gama
Eficiência de detetores Ge(Li)
Resumen en portugués
Foi efetuado o mapeamento da eficiência intrínseca no fotopico de dois detetores Ge(Li) do tipo falso coaxial, com fótons colimados de uma fonte de 152Eu de 1 mCi, em f unção da direção de incidência. Este mapeamento permitiu efetuar experimentalmente a correção de ângulo sólido para experimentos de correlação angular gama-gama. Esta correção não possui hipóteses como: detetor perfeitamente cilíndrico, ou com seu eixo coincidente com aquele do invólucro. A aplicação da informação de eficiência para a correção de dados de correlação angular implicou em um novo método de análise, onde é ajustada a eles uma expressão do tipo: W(?)= ?.[1+ ?_(kpar>0)??Akk .?_(m= -k)^k?C_km .e^(-1m?)]? onde os Ckm são coeficientes complexos. A qualidade do método foi verificada com a análise de correlações angulares presentes nos decaimentos dos isótopos 60Co e 100Rh com misturas multipolares conhecidas. Os resultados obtidos foram satisfatórios, demonstrando a qualidade dos métodos de mapeamento e de análise da função de correlação angular.
Título en inglés
Solid angle experimental correction for gamma-gamma angular correlation
Palabras clave en inglés
Gamma-gamma angular correlation
Ge (Li) detectors efficiency
Solid angle correction
Resumen en inglés
The scanning of the intrinsic photopeak efficiency for two Ge(Li) detectors, of false coaxial type (wrap around) was performed with collimated photons from a 152Eu source with 1 mCi, as a function of the direction of incidence. This scanning allowed to obtain experimentally the solid angle correction, to be used in gamma-gamma angular correlation experiments. This correction have no hypotheses like: detector perfectly cylindrical, or with its axis coincident with the end cup axis. The efficiency information applied for the correction of angular correlation data involved a new analysis method, where it is fitted to the data an expression Iike this W(?)= ?.[1+ ?_(kpar>0)??Akk .?_(m= -k)^k?C_km .e^(-1m?)]? the Ckm are complex coefficients. The quality of the method was checked with data analysis of angular correlation functions present in the 60Co and 100Rh decays, with known multipole mixtures. The results were satisfactory, demonstrating the quality of both methods, detector mapping and analysis of correlation functions.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Procida_1990.pdf (4.10 Mbytes)
Fecha de Publicación
2015-06-17
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.