Master's Dissertation
DOI
https://doi.org/10.11606/D.43.1990.tde-07052014-181341
Document
Author
Full name
Eduardo Luiz Augusto Macchione
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 1990
Supervisor
Committee
Dietzsch, Olacio (President)
Cusatis, Cesar
Ueta, Nobuko
Title in Portuguese
Detector de raios-X sensível à posição.
Keywords in Portuguese
Detector proporcional sensível à posição
Instrumentação nuclear
Raios-x
Abstract in Portuguese
Um contador a gás, com multifios, sensível à posição, para a detecção de raios X foi construído utilizando linhas de atraso integradas para leitura da posição. Seis linhas de atraso (50ns de atraso cada, frequência de corte de 40MHz) cobrem um comprimento sensível total de 150mm, possuindo uma razão do atraso total pelo tempo de subida dos sinais na linha adequadas para a detecção em posição raios X de 5,9keV do ANTIPOT. 55 Fe com alta resolução. Testes usando a forneceram uma não-linearidade linha de integral máxima menor que 0,1% e não-linearidade diferencial máxima de ±4,0%, com o detector utilizando a mistura 700torr. Testes similares foram feitos usando de Ar-CH IND 4 (90%-10%) a a linha de 8,04keV de um tubo de raios X de Cu. Uma resolução total de 330m, e as mesmas não-linearidades integral e diferencial foram obtidas. Com a finalidade de t.est.ar o comportamento do detector sob condições experimentais reais, duas diferentes amostras foram analisadas na região de espalhamento em baixo ângulo. As amostras analisadas foram o carvão vítreo e o dodecil sulfato de sódio (SLS). O feixe de raios X foi colimado por um conjunto de fendas, com 50,um cada uma delas e separadas de 98mm. A distância entre a amostra e o detector era de 404,5mm, fazendo com que a região sensível do detector cobrisse um ângulo de espalhamento de ±12 °. Foram obtidos um raio de giro de 22 Å para o carvão vítreo e um dos parâmetros da rede cristalina c de 39,1 Å para o SLS. Estes resultados estão de acordo com trabalhos anteriores obtidos com técnicas tradicionais de maior dispêndio de tempo que utilizam detectores de Nal(T1) e chapas fotográficas em uma câmara de Laue.
Title in English
X-rays detector sensitive to the position.
Keywords in English
Nuclear Instrumentation
proportional detector sensitive to the position
X-rays
Abstract in English
A multiware position sensitive gas counter for X-Ray detection was developed in our laboratory making use or commercial delay-lines for position sensing. Six delay-line chips (50ns delay each 40Mhz cut-off frequency) cover a total sensitive length of 190mm leading to a delay-risetime ratio that allows for a high-resolution position detection. Tests using the 5,9keV X-Ray line from a 55Fe source an integral linearity better than 0,1% and a maximum differential linearity or ±4,0% were obtained operating the detector with an Ar-CH 4 (90%-10%) gas mixture at 700torr. Similar tests were performed using the 8,04keV line from a Cu X-Ray tube. A total resolution or 330m, and the same integral and differential linearities were obtained. In order to check the detector performance under usual experimental conditions two different samples were analyzed in the low scattering-angle region. The samples analyzed were vitreous carbon and sodium dodecyl sulfate (SLS). The X-Rays were collimated by a set or slits each 50m wide separated by 98mm. The distance between the sample and the detector was 404,5mm, so that the detector covered a scattering angle of ±12°. A radius of gyration of 22Å for vitreous carbon and a crystallographic parameter of 39,1Å for SLS were obtained. These values are in good agreement with those obtained with the more time-consuming traditional techniques making use of a rotating NaI(T1) detector and photographic plates with a Laue camera.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Macchione.pdf (7.24 Mbytes)
Publishing Date
2014-05-16
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.