Thèse de Doctorat
DOI
https://doi.org/10.11606/T.43.1998.tde-04122013-145919
Document
Auteur
Nom complet
Eduardo Luiz Augusto Macchione
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 1998
Directeur
Jury
Dietzsch, Olacio (Président)
Cruz, Manoel Tiago Freitas da
Souza, Gerardo Gerson Bezerra de
Ueta, Nobuko
Vieira Junior, Nilson Dias
Titre en portugais
Espectrômetro de Massa por Tempo de Vôo para Análise de Macromoléculas
Mots-clés en portugais
Física nuclear
Instrumentação
Resumé en portugais
Um espectrômetro de massa por tempo de vôo foi projetado e construído. A técnica de Plasma Desorption é empregada para provocar a dessorção e ionização das moléculas da amostra por colisões com os produtos de fissão emitidos de uma fonte radioativa de ANTPOT.252 Cf. O espectrômetro encontra-se instalado no interior de uma câmara de vácuo, que permite a sua operação a pressões da ordem de 10 POT.-7 torr. Uma pré-câmara é utilizada para a manipulação e troca de até três amostras simultaneamente, sem a necessidade da abertura da câmara principal para a atmosfera, reduzindo o tempo de recuperação da pressão do sistema para início das análises. No tubo de vôo, que possui uma extensão de 111 cm, encontra-se instalado um guia eletrostático de partículas, usado para aumentar a transmissão dos íons secundários até o sistema de detecção. Medidas foram feitas utilizando-se amostras-padrão para a caracterização do instrumento. Estudos também foram realizados sobre o comportamento da resolução e da eficiência de transmissão do aparelho em função entre 1600 e 1800 para íons com massas em torno de 900 u.m.a.
Titre en anglais
Time-of-flight mass spectrometer for analysis of macromolecules
Mots-clés en anglais
Instrumentation
Nuclear physics
Resumé en anglais
A time-of-flight mass spectrometer was projected and constructed. The Plasma Desorption technique is employed to desorb and ionize the molecules sample by collisions with fission fragments emitted from ANTPOT.252 Cf source. The spectrometer is placed inside a vacuum chamber, allowing operation at a pressure around of 10 POT.-7 torr. A pre-chamber is used to insert three samples simultaneously into the mass spectrometer, without the need of opening the main chamber to the atmosphere, reducing the recover time of the pressure system for the beginning of analyzes. Inside the flight tube (111 cm length) was installed an electrostatic particle guide, used to increase the secondary ion transmission to the detection system. Measurements using standard samples were done to characterize the apparatus performance. The instrument resolution and transmission efficiency were analyzed with respect to the secondary ion acceleration voltage. The mass resolution of m/m from 1600 to 1800 for rnass ions around 900 u.m.a. was obtained.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2014-02-20
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.