Disertación de Maestría
DOI
https://doi.org/10.11606/D.25.2013.tde-14102013-164847
Documento
Autor
Nombre completo
Aline Baia Miranda
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Bauru, 2013
Director
Tribunal
Araujo, Carlos dos Reis Pereira de (Presidente)
Conti, Paulo Cesar Rodrigues
Thomé, Geninho
Título en portugués
Análise histomorfométrica dos implantes de titânio grau 4 e grau 5: estudo experimental em coelhos
Palabras clave en portugués
Liga de titânio
Regeneração óssea
Titânio
Resumen en portugués
Os implantes dentários de titânio comercialmente puro (Ticp) grau 4 e os de titânio-alumínio-vanádio (Ti6Al4V) grau 5, possuem boas propriedades mecânicas. No entanto, algumas situações clínicas com restrição de espessura óssea impedem a utilização dos implantes com 3.75mm de diâmetro tradicionalmente fabricados em titânio 4 e que apresentam resistência mecânica testada em profusão. Como a liga de grau 5 possui superior resistência à tração e à fadiga, este estudo objetiva analisar a resposta óssea do titânio grau 4 e grau 5 em implantes curtos e estreitos através da análise do contato osso-implante (BIC) e da área de neoformação óssea. Para este fim, o presente estudo utilizou 15 coelhos da raça New Zealand, que receberam um total de trinta implantes divididos em suas tíbias direita e esquerda. Os implantes de grau 4 e de grau 5, com dimensões de 3.5x8mm e 2.9x7mm, respectivamente, foram fornecidos pela empresa NEODENT® (Curitiba-Brasil). Neste estudo, dois grupos foram formados, um com implantes de titânio grau 4 e outro com implantes de titânio grau 5, ambos contendo quinze implantes curtos e estreitos. Cortes histológicos foram realizados após duas semanas de osseointegração. Mensurações no analisador de imagens ImageJ foram feitas para verificar o BIC e a área de osso neoformado. Para a análise dos dados estatísticos, utilizou-se o teste t de Student para amostras independentes com nível de significância de p<0.05. Os resultados mostraram que não houve diferença estatisticamente significante entre os grupos nos dois tipos de análises. A média dos valores de BIC do titânio grau 4 obteve o valor médio mais alto de 56,53%, variando de 95,74% a 9,4%. O grupo de titânio grau 5 apresentou média de 50,63%, variando entre 84,3% e 13,12%. Com base em imagens com fluorescência, realizou-se análise da área óssea. O grupo de titânio grau 5 apresentou valor de 46,3% (variando entre 79% a 14,21%) em fluorescência, número ligeiramente maior do que o verificado no grupo de titânio grau 4 (44,73%, com variância entre 78,93% e 18,36%). Com isso, concluiu-se que implantes de titânio de grau 4 e de grau 5 obtiveram respostas biológicas equivalentes, quando avaliados o BIC e a área óssea neoformada.
Título en inglés
Histomorphometric evaluation of titanium implants grade 4 and grade 5: experimental study in rabbits
Palabras clave en inglés
Bone regeneration
Titanium
Titanium alloy
Resumen en inglés
Commercially pure titanium implants (Ticp) of degree IV and of titanium-aluminum-vanadium degree V have good mechanical properties. However some clinical situations with limited bone thickness preclude the use of traditionally implants with 3.75mm diameter made from titanium IV and present strength tested in profusion. In addition, degree V alloy shows higher resistance to traction and fatigue, this study aims to analyze the bone response to degree IV and degree V titanium in short and narrow implants with the same surface treatment through the analysis of the bone-to-implant contact (BIC) and the area of the new bone formation. For such purpose, this study used15 New Zealand white rabbits which had a total of thirty implants inserted in their right and left tibiae. Both degree IV and degree V implants, measuring 3.5x8mm and 2.9x7mm, respectively, were provided by dental implant manufacturer NEODENT® (Curitiba-Brazil). The implants were allocated into two distinct groups. Fifteen short and narrow implants of degree IV were utilized in a group, whereas fifteen short and narrow implants of degree V were employed in the other group. Histological sections samples were carried out after a two-week period of osseointegration. Subsequently, measurements were calculated using ImageJ software in order to verify the BIC and the area of the newly grown bone. Analysis of statistical data of independent samples with significance level p<0.05 was carried out via Student's t-test. Results showed no statistically substantial difference between both groups regarding both measurement activities. The media values BIC for the degree IV titanium group reached the highest average value of 56,53%, varying from 95,74% to 9,4%. The degree V titanium group averaged 50,63%, oscillating between 84,3% and 13,12%. The bone areas were analyzed based on images with fluorescence. The degree V titanium group reached the percentage of 46,3% (varying from 79% to 14,21%) in fluorescence whereas the degree IV titanium group had a percentage of 44,73% (varying between 78,93% and 18,36%). Therefore, both degree IV and degree V implants had equivalent biological responses based on the analysis of BIC and newly grown bone areas.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2013-10-16
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.