Disertación de Maestría
DOI
https://doi.org/10.11606/D.25.2005.tde-28032006-145556
Documento
Autor
Nombre completo
Marcelo Junior Zanda
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Bauru, 2005
Director
Tribunal
Capelozza, Ana Lucia Alvares (Presidente)
Carvalho, Izabel Maria Marchi de
Haiter Neto, Francisco
Título en portugués
Desenvolvimento de um posicionador para padronização da tomada radiográfica oclusal
Palabras clave en portugués
estomatologia
maxila
oclusão dentária
Resumen en portugués
Visando padronizar e simplificar a técnica radiográfica oclusal total de maxila foi desenvolvido um Posicionador em acrílico para filmes radiográficos oclusais. O dispositivo reproduz as especificações da técnica radiográfica oclusal total de maxila: distância focal de 40 cm, ângulo vertical de 65°, ângulo horizontal de 0º e feixe central de raios X incidindo na glabela. A eficácia deste dispositivo foi avaliada a partir de radiografias oclusais totais de maxila realizadas por três operadores com o Posicionador em treze crânios humanos e comparadas com a técnica radiográfica convencional (sem o uso do posicionador) realizada pelos mesmos operadores. Sete pontos anatômicos de fácil visualização foram marcados nas radiografias para a determinação de medidas lineares entre estes pontos. Estas medidas serviram para a comparação dos resultados dos dois grupos: Grupo 1 (radiografias executadas com o Posicionador) e Grupo 2 (sem o Posicionador). Os dados obtidos foram tratados estatisticamente pelo teste de erro de Dahlberg, Análise de Variância a Dois Critérios, teste de Tukey e teste "t" pareado nas médias dos desvios padrão de cada operador. Os resultados demonstraram a viabilidade de uso do Posicionador idealizado e maior padronização dos resultados nas radiografias oclusais totais de maxila realizadas com este dispositivo.
Resumen en inglés
The purpose of the present study was to project and construct a Film-Holding Instrument to obtain occlusal radiographs. The positioner was made of acrylic and geometric calculations were done to determine the correct technical specification: 40 cm (16 inch) focal length, 65º vertical angle, 0º horizontal angle, film parallel to the occlusal plane and primary beam of the x-ray focusing the bridge of the nose. Three examiners performed 78 radiographs in 13 human skulls, 39 radiographs were obtained using the positioning and 39 radiographs were obtained without the positioner. Seven anatomic structures were marked on radiographs and used as references for linear measurements. All measurements were submitted to statistics analyze: Two Way Anova Tukey test and "T" test. Results demonstrated that the new Film-Holding Instrument was effective.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
MarceloJuniorZanda.pdf (475.70 Kbytes)
Fecha de Publicación
2006-04-06
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.