Disertación de Maestría
DOI
https://doi.org/10.11606/D.23.2012.tde-16012013-105147
Documento
Autor
Nombre completo
Piero Rocha Zanardi
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2012
Director
Tribunal
Mori, Matsuyoshi (Presidente)
Pereira Filho, Júlio
Sesma, Newton
Título en portugués
Intercambialidade de componentes protéticos em diferentes marcas de implantes
Palabras clave en portugués
Adaptação protética
Implante Dentário
Microscopia Eletrônica de Varredura
Pilar Protético
Resumen en portugués
O presente estudo avaliou a intercambiabilidade dos componentes protéticos de implantes de hexágono externo medindo-se a precisão da interface implante/pilar (I/P) com microscopia eletrônica de varredura. Foram utilizados dez implantes para cada uma de três marcas (SIN, Conexão, Neodent) com seus respectivos pilares (base metálica de CoCr , rotacional e não rotacional) e um de marca alternativa (Microplant) em um arranjo com todas as combinações de I/P possíveis. O valor de referência para a intercambiabilidade das várias marcas de componentes foi definido pela diferença do pilar original para com seu respectivo implante. Dessa maneira, a intercambiabilidade foi considerada válida quando o resultado para um determinado pilar colocado sobre um implante fosse igual ou menor que a diferença medida, quando o pilar da mesma marca do implante foi posicionado. Um grau de desajuste foi observado em todos os espécimes. De uma maneira geral, o pilar antirrotacional foi mais preciso do que o rotacional. Estes variaram de 0,6 16,9m, com mediana 4,6m, enquanto o antirrotacional foi de 0,3 12,9m, com mediana de 3,4m. Os espécimes com o pilar e o implante Conexão tiveram uma diferença maior que o conjunto original para SIN e NEO (p<0,05). Apesar de estes últimos apresentarem resultados semelhantes com seus respectivos pilares, quando trocados, não mantiveram a mesma precisão original. Os resultados sugerem que a marca alternativa seria a única que apresentaria compatibilidade com todos os sistemas, enquanto as outras marcas não seriam completamente intercambiáveis.
Título en inglés
Evaluation of the connecting accuracy of interchanged prosthetic abutments to different dental implants using scanning electron microscopy
Palabras clave en inglés
Dental abutments
Dental implants
Dental prosthesis design
Microscopy electron scanning
Prosthesis fitting
Resumen en inglés
The present study evaluated the interchangeability of prosthetic components for external hexagon implants by measuring the precision of the implant/abutment (I/A) interface using scanning electron microscopy. Ten implants for each of three brands (SIN, Conexão, Neodent) were tested with their respective abutments (milled CoCr collar rotational and non-rotational) and that of an alternative manufacturer (Microplant) in randomly arranged I/A combinations. The degree of interchangeability between the various brands of components was defined using the original abutment interface gap with its respective implant as the benchmark dimension. Accordingly, when the result for a given component placed on an implant was equal to or smaller then that gap measured when the component of the same brand as the implant was positioned, interchangeability was considered valid. Some degree of maladjustment was observed on all specimens. Generally, the non-rotational component was more accurate than its rotational counterpart. The latter samples ranged from 0.6 16.9m, with a median of 4.6m; the former from 0.3 12.9m, with a median of 3.4m. Specimens with the abutment and fixture from Conexão had larger microgap than the original set for Sin and Neodent (p<0.05). Even though the latter systems had similar results with their respective components, their interchanged abutments did not reproduce the original accuracy. The results suggest that the alternative brand abutment would have compatibility with all systems while the other brands were not completely interchangeable.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2013-03-14
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.