Dissertação de Mestrado
DOI
https://doi.org/10.11606/D.23.2008.tde-20012009-102437
Documento
Autor
Nome completo
Luis Cesar Brisighello
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2008
Orientador
Banca examinadora
Gavini, Giulio (Presidente)
Caldeira, Celso Luiz
Duarte, Marco Antonio Hungaro
Título em português
Resistência à fadiga flexural dos sistemas rotatórios k3 e Endosequence em razão do uso
Palavras-chave em português
Fadiga cíclica
Fratura
Níquel-titânio
Resumo em português
O presente estudo teve como objetivo avaliar a resistência à fadiga cíclica de dois diferentes sistemas rotatórios de níquel-titânio, K3 (Sybron Endo, EUA) e Endosequence (Brasseler, EUA), fundamentando-se no número de usos. Todas as limas rotatórias que foram selecionadas possuíam conicidade 0.04, 25 mm de comprimento e diâmetro de ponta 25, padrão ISO. Para tal logro, um dispositivo desenvolvido especificamente para executar ensaios dinâmicos foi utilizado. Cada grupo foi subdividido em quatro subgrupos em função do número de usos, sendo estabelecidos da seguinte maneira: grupos A0 e B0, instrumentos sem nenhum uso; grupos A1 e B1, instrumentos de um único uso; grupos A3 e B3, instrumentos de três usos e grupos A5 e B5, instrumentos de cinco usos. Cada subgrupo era composto por 12 limas, totalizando 96 instrumentos rotatórios entre K3 e Endosequence. A simulação foi realizada em canais artificiais de resina com curvatura de 40 graus e raio de 5 mm. Todas as limas foram submetidas a ensaios de fadiga cíclica realizados em um dispositivo experimental que permitiu que o instrumento reproduzisse uma instrumentação rotatória em canais curvos. Esse dispositivo possui um temporizador, que registra o tempo de avanços em segundos, desde o início do movimento até a fratura da lima e também um contador, que registra dentro do intervalo de tempo o número de ciclos realizados pelo cilindro pneumático até o momento da fratura. Foi utilizada uma peça de mão com contra-ângulo redutor de 16:1, acionado por motor elétrico na velocidade de 350 rpm e 2 Ncm de torque. O tempo foi registrado por meio do contador presente no dispositivo. Os valores foram transformados em segundos. Para análise estatística, empregou-se o teste de análise de variância (dois critérios) entre as amostras testadas e observou-se que não houve diferença estatisticamente significante em relação ao número de usos. Entretanto, o sistema K3 apresentou maior resistência à fadiga flexural em relação ao sistema Endosequence (p < 0,05).
Título em inglês
Resistance to flexural fatigue of K3 and Endosequence rotary systems in relation to the use
Palavras-chave em inglês
Cyclic fatigue
Fracture
Nickel-titanium
Resumo em inglês
The aim of this work was to evaluate the cyclic fatigue resistance of two different nickel-titanium rotary systems, K3 (Sybron Endo, EUA) and Endosequence (Brasseler, EUA), based on the number of uses. All rotary files which were selected had their conicity of 0,04, 25mm of length and tip diameter of 25, following ISO standardization. To reach the purposes, a specially developed apparatus to perform dynamic assays was employed. Each group was subdivided into 4 subgroups according to the number of uses, being established as follows: groups A0 and B0, instruments without any use; groups A1 and B1, instruments of a single use; groups A3 and B3, instruments of three uses and groups A5 and B5, instruments of five uses. Each subgroup was compound of 12 files, totalizing 96 rotary instruments between K3 and Endosequence. The simulation was conducted in resin artificial canals presenting curvature of 40 degrees and radius of 5 mm. All files were submitted to cyclic fatigue assays performed on an experimental apparatus where the instrument could reproduce a rotary instrumentation in curved canals. This equipment contains an timer, which records advance time per seconds, since the begin of movement until the file fracture, and also has a counter which registers, on a period of time, the cycle number accomplished by the pneumatic cylinder until the fracture time. A reducer contra-angle hand piece of 16:1 was used, activated by an electric motor on a speed of 350rpm and 2 Ncm of torque. Time was recorded by the counter coupled in the apparatus. Values were converted into seconds. The two-way analysis of variance was employed to statistical evaluation, among tested samples and we could observe that no statistical differences were found in relation to the number of uses. However, the K3 system showed higher resistance to flexural fatigue when compared to the Endosequence system (p < 0,01).
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
LUISCESARBRISIGHELLO.pdf (594.07 Kbytes)
Data de Publicação
2009-03-16
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.