Master's Dissertation
DOI
https://doi.org/10.11606/D.18.2003.tde-28102015-151149
Document
Author
Full name
José Cláudio Pinto de Azevedo
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 2003
Supervisor
Committee
Mucheroni, Mario Francisco (President)
Di Giacomo, Benedito
Tsunaki, Roberto Hideaki
Title in Portuguese
Calibração de acelerômetros
Keywords in Portuguese
Calibração
Interferometria laser
Rastreabilidade
Abstract in Portuguese
Este trabalho apresenta o projeto e desenvolvimento de dois sistemas de calibração de acelerômetros. O primeiro sistema foi desenvolvido a partir de um sistema interferométrico laser Hewlett-Packard (HP 5529A), utilizado como padrão de referência absoluta. Este sistema foi projetado com o propósito de calibrar acelerômetros-padrão de comparação e acelerômetros-padrão de transferência. O segundo sistema de calibração utiliza como padrão de referência acelerômetros-padrão de comparação. Este segundo sistema foi projetado com o objetivo de calibrar acelerômetros de uso corriqueiro. Desta forma, o conjunto formado pelos dois sistemas de calibração podem ser rastreáveis até os padrões primários, nacionais e internacionais. Para cada um dos sistemas de calibração são apresentadas discussões relativas às suas características construtivas, aos testes de operação e às avaliações sobre os desempenhos alcançados.
Title in English
Calibration of accelerometers
Keywords in English
Calibration
Laser interferometric system
Traceability
Abstract in English
This work presents the design and development of two accelerometer calibration systems. The first system was developed from a Hewlett-Packard laser interferometric system (HP 5529A), taken as an absolute reference standard. This system was designed to allow calibration of comparison standard accelerometers and transfer standard accelerometers. The second calibration system, which employs comparison standard accelerometers as reference, was designed to calibrate common use accelerometers. Thus, the arrangement of the two calibration systems can be traceable until the primary national and international standards. For each calibration system, a discussion is presented, concerning their constructive characteristics, operation testing and evaluation of accomplished performance.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2015-10-28
 