• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Dissertação de Mestrado
DOI
https://doi.org/10.11606/D.17.2017.tde-04012017-155803
Documento
Autor
Nome completo
Leandro Federiche Borges
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
Ribeirão Preto, 2016
Orientador
Banca examinadora
Oliveira, Harley Francisco de (Presidente)
Marques, Paulo Mazzoncini de Azevedo
Marta, Gustavo Nader
Pavoni, Juliana Fernandes
 
Título em português
Validação da metodologia de controle da qualidade in vivo com auxílio de filme radiocrômico, aplicados à teleterapia
Palavras-chave em português
Dosimetria
Filme radiocrômico
IMRT
Radioterapia
Resumo em português
A aferição da dose por técnica de dosimetria in vivo garante que a dose calculada pelo sistema de planejamento seja àquela entregue pelo acelerador linear (AL) durante a radioterapia, garantindo assim os processos de controle de qualidade. O objetivo deste trabalho foi validar a técnica de controle de qualidade em IMRT utilizando filme radiocrômico. Foram analisados 47 planejamentos de IMRT de três regiões: próstata, cabeça e pescoço e crânio. Para cada planejamento utilizouse um filme radiocrômico EBT2 acoplado ao cabeçote do acelerador linear através de uma bandeja preparada para o filme. A leitura e posterior análise da dose pontual e distribuição de dose após 24 horas, foi realizada em sofware desenvolvido em MatLab. O índice de aprovação foi de 92%. Quatro planos foram reprovados em pelo menos 1 critério, sendo 3 de cabeça e pescoço e 1 de crânio. Os resultados da dosimetria in vivo com filme radiocrômico foi validada, podendo ser considerada uma técnica confiável e prática na aferição da dose em radioterapia.
 
Título em inglês
Validation of in vivo quality control methods with the aid of radiochromic film, applied to teletherapy
Palavras-chave em inglês
Dosimetry
IMRT
Radiochromic film
Radiotherapy
Resumo em inglês
The measurement of dose, using the in vivo technique dosimetry, ensures that the dose calculated by the planning system is exactly the same as the dose delivered in linear accelerator, this ensures quality assurance in radiotherapy. The objective of this study was to evaluate the quality control techinique of IMRT treatments using radiochromic film. We analyzed 47 IMRT plans, which divided into three regions: Prostate, head and neck, and skull. For each plan, we used a radiochromic film coupled in linear accelerator head by a tray prepared to film placement. The film was stored for 24 hours after irradiation. Reading and analysis of point dose and 2D dose distribution were performed after this period. ll plans analyzed, the approval rating was 92%, and the others were flunked at least one criterion. Based on the results obtained in vivo dosimetry with radiochromic film was validated. This technique is reliable and practical to measure the dose delivered by linear accelerator.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2017-03-28
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores.
CeTI-SC/STI
© 2001-2024. Biblioteca Digital de Teses e Dissertações da USP.