Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.17.2014.tde-24062014-084525
Document
Auteur
Nom complet
Juliana dos Santos Ribeiro
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
Ribeirão Preto, 2014
Directeur
Jury
Ferreira, Cristine Homsi Jorge (Président)
Brito, Luiz Gustavo Oliveira
Oliveira, Andrea Lemos Bezerra de
Titre en portugais
Avaliação da reprodutibilidade de medidas obtidas com o uso de perineômetro no período gestacional
Mots-clés en portugais
Assoalho pélvico
Gestação
Reprodutibilidade dos resultados
Resumé en portugais
O objetivo primário deste estudo foi avaliar a reprodutibilidade inter-examinador, na mensuração da pressão intravaginal em gestantes. Participaram do estudo 41 gestantes baixo risco com capacidade de contrair voluntariamente os MAP. Os MAP (músculos do assoalho pélvico) foram avaliados quanto a sua capacidade de contração por meio da palpação vaginal , e em seguida com uso do perineômetro Peritron®. Foram realizadas três contrações máximas com intervalo de descanso entre cada contração de 10 segundos. As avaliações foram realizadas por dois examinadores distintos e cegos em relação aos resultado da avaliação do outro, em dois dias alternados, com intervalo entre as avaliações de 2 a 7 dias, mantendo o mesmo horário entre as avaliações. As gestantes tinham idade entre 18 e 41 anos (26,07±5,69), índice de massa corpórea (IMC) entre 18,32 e 43,43 Kg/m² (26,46±5,77), com paridade entre 0 e 8 partos (1,09±1,65). Na análise estatística foi utilizado Coeficiente de Correlação de Concordância (CCC) para as medidas entre avaliadores e os limites de concordância de Bland e Altman foram utilizados para observar a concordância. Para medir a reprodutibilidade intra-sessão das três medidas para cada avaliador foi utilizado o Coeficiente de Correlação Intraclasse (ICC). Os resultados foram interpretados de acordo com a seguinte classificação: valores abaixo de 0,50 foram considerados de reprodutibilidade pobre, valores 0,50 a 0,75 foram considerados como moderada reprodutibilidade e valores acima de 0,75 foram considerados de boa reprodutibilidade. Foram encontradas boa concordância nas medidas interexaminadores entre os valores da média do pico no período gestacional (0,81) e essa concordância também ocorreu quando os trimestres foram analisados distintamente (0,81, 0,78, 0,86 para primeiro, segundo e terceiro trimestre respectivamente). Quanto a variável pico de contração os resultados mostraram uma concordância que variou de moderada à boa, para a análise das aferições separadamente, com valores que variaram de 0,54 à 0,88. Já as variáveis média e duração da contração apresentaram valores irregulares. A reprodutibilidade das medidas intra-sessão para a variável pico, para o primeiro, segundo, terceiro trimestre gestacional e para o período gestacional geral mostraram valores de concordância moderada para o avaliador 1 (0,72, 0,75, 0,78, 0,74 respectivamente) e valores de forte concordância para o avaliador 2 (0,95, 0,93, 0,97, 0,94 respectivamente). O presente estudo concluiu que o perineômetro é um método com boa reprodutibilidade inter-examinador para avaliar os músculos do assoalho pélvico (MAP) durante o período gestacional, quando as variáveis de média do pico da contração e pico da contração são utilizadas. Esse estudo não encontrou diferença entre os trimestres gestacionais, sugerindo que a reprodutibilidade da média do pico de contração e pico de contração do AP utilizando-se o perineometro Peritron é variável de moderada a boa.
Titre en anglais
Evaluation of the reproducibility of measurements obtained with the use of perineometer during pregnancy
Mots-clés en anglais
Pelvic Floor
Pregnancy
Reproducibility of results
Resumé en anglais
The primary objective of this study was to evaluate the inter-examiner reproducibility in the measurement of intravaginal pressure in pregnant women. The study included 41 low-risk pregnant women. The ability to contract PFM (pelvic floor muscles) was assessed using vaginal palpation and the PFM strength was measured using the perineometer Peritron. Three maximal contractio ns were performed with an interval of rest between each contraction of 10 seconds. The evaluations were performed by two different examiners who were blinded to each evaluation result of the other, on two alternate days, with an interval between assessment s 2-7 days keeping the same time between assessments. The women were aged between 18 and 41 years (26.07 ± 5.69), body mass index (BMI) between 18.32 and 43.43 kg / m² (26.46 ± 5.77), with parity 0 births and 8 (1.09 ± 1.65). For statistical analysis we used the concordance correlation coefficient (CCC) for measurements between examiners. Bland and Altman limits of agreement was used in the analysis. To measure the intra-session reproducibility of the three measures for each examiner the intraclass correlation coefficient (ICC) was used. The results were interpreted according to the following classification: values below 0.50 were considered poor reproducibility, values from 0.50 to 0.75 were considered moderate reproducibility and values above 0.75 were considered good reproducibility. It was found a good concordance of inter-examiner between the values of mean peak during pregnancy (0.81) and this correlation also occurred when the gestational trimesters were analyzed distinctly (0.81, 0.78, 0.86 for first, second and third trimester respectively). As for variable peak contraction results showed a concordance ranging from moderate to good, for the analysis of the measurements separately, with values ranging from 0.54 to 0.88. The average and durati on of contractions showed irregular values . The reproducibility of intra-session measures for the variable peak, for the first, second, and third trimester of pregnancy for general values showed moderate agreement for evaluator 1 (0.72, 0.75, 0.78, 0 , 74 respectively) and values of concordance for evaluator 2 (0.95, 0.93, 0.97, 0.94 respectively). This study concluded that perineometer is a method with good reproducibility inter -examiner to evaluate the pelvic floor muscles during pregnancy, when th e variables of mean peak contraction and peak contraction are used. This study found no difference between gestational trimesters, suggesting that interrater reproducibility of of mean peak contraction and peak contraction of PFM using Peritron perineomet er is variable from moderate to good.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2015-06-08
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.