Doctoral Thesis
DOI
https://doi.org/10.11606/T.17.2020.tde-20082020-113301
Document
Author
Full name
Gustavo Henrique Vieira de Andrade
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
Ribeirão Preto, 2019
Supervisor
Committee
Abud, Daniel Giansante (President)
Garcia, Rodrigo Gobbo
Muglia, Valdair Francisco
Tucci Junior, Silvio
Title in Portuguese
Exposição à radiação ionizante dos pacientes e operador na embolização das artérias prostáticas para tratamento da hiperplasia benigna
Keywords in Portuguese
Dose absorvida
Embolização terapêutica
Equivalente de dose pessoal
Hiperplasia prostática
Máxima dose na pele
Proteção radiológica
Sintomas do trato urinário inferior
Abstract in Portuguese
A embolização das artérias prostáticas (EAP) para tratamento dos sintomas obstrutivos da hiperplasia benigna é um procedimento complexo pela variabilidade anatômica, superposição de vasos e estruturas pélvicas, além do fino calibre e tortuosidade destas artérias. Embora venha ganhando cada vez mais aceitação, não tem sido alvo de estudos dosimétricos. Este trabalho apresenta os resultados de exposição à radiação ionizante dos pacientes e do operador em 34 pacientes consecutivos submetidos à EAP no Hospital da Restauração, Recife-PE. Dados epidemiológicos gerais e do procedimento foram obtidos tais como tempo de fluoroscopia, parâmetros de radiação, projeções e número de imagens. A dosimetria dos pacientes foi realizada através da dose máxima na pele (PSD) obtida com filme radiocrômico (Gafchromic XR RV3), pelo produto kerma ar-área (PKA) e do kerma ar no ponto de referência (Ka,r). A dosimetria ocupacional foi realizada com dosímetros termoluminescentes distribuídos em várias regiões do radiologista intervencionista. As primeiras 25 EAPs foram realizadas com protocolo padrão do equipamento. Neste grupo, a média da máxima dose na pele dos pacientes foi de 2,4 Gy, com PKA médio de 450.7 Gy.cm2. A dose efetiva média do médico foi de 17μSv. Após implementação e testes em phantoms do protocolo otimizado de radiação denominado RECiFE (Radiation Exposure Curtailment For Embolization) desenvolvido em parceria e baseado no sistema CARE - Siemens (Combined applications to reduce exposure), foram realizadas 9 EAPs e os resultados dos protocolos foram comparados. Houve redução das doses do operador de aproximadamente 80%. Concluímos que a EAP é um procedimento com elevada exposição dos pacientes e operador e a otimização dos parâmetros é mandatória, questionando-se os protocolos convencionais dos fabricantes.
Title in English
Radiation exposure of patients and staff in prostatic artery embolization for benign hyperplasia
Keywords in English
Absorbed dose
Embolization
Lower urinary tract symptoms
Maximum skin dose
Personal equivalent dose
Prostatic hyperplasia
Radiological protection
Abstract in English
Prostatic artery embolization (PAE) for treatment of obstructive symptoms caused by benign hyperplasia is a technically challenging procedure due the complex anatomic variations, superimposed pelvic vessels and structures and thin and tortuous prostatic arteries. Although its increasingly acceptance, very few dosimetry studies have been done. This paper presents the radiation exposure results of patients and operator in 34 consecutive PAEs in Hospital da Restauração, Recife-PE. Demographic and general procedures data were obtained such as fluoroscopy time, irradiation parameters, used projections and number of images. Patient dosimetry comprised the measure of peak skin dose (PSD) with radiochromic films (Gafchromic XR RV3), air kerma-area product (PKA) and reference air kerma (Ka,r). Occupational dosimetry was performed using thermoluminescent dosimeters distributed over various regions of the physician's body. The first 25 PAEs were performed using the standard radiation protocol of the equipment. In this group the mean PSD was 2,4 Gy, and the mean PKA 450,7 Gy.cm2. The operator mean effective dose was 17μSv. After implementation and phantom testing of optimized protocol called RECiFE (Radiation Exposure Curtailment For Embolization) based on Siemens CARE (Combined applications to reduce exposure), we performed 9 PAEs and compared the exposure parameters with standard protocol. A reduction of approximately 80% was obtained in staff doses. Thus, PAE is high exposure procedure for patients and operator. Optimization is mandatory and standard equipment protocol must be examined.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2020-10-19
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.