Disertación de Maestría
DOI
https://doi.org/10.11606/D.9.2008.tde-16062008-222314
Documento
Autor
Nombre completo
Maurício Rocha de Magalhães Sampaio
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2008
Director
Tribunal
Porta, Valentina (Presidente)
Costa, Maria Teresa Zulini da
Santos, Silvia Regina Cavani Jorge
Título en portugués
Avaliação da bioequivalência de formulações contendo lorazepam através de método bioanalítico utilizando a cromatografia líquida acoplada ao sistema de detecção por espectrometria de massa
Palabras clave en portugués
Biodisponibilidade (Avaliação)
Bioequivalência (Avaliação)
Bromazepam
Fármacos psicoterápicos (Avaliação; Análise quantitativa)
LC-MS/MS
Lorazepam
Plasma humano
Resumen en portugués
Desenvolveu-se método de alta sensibilidade e especificidade por cromatografia liquida de alta eficiência acoplada a detecção por espectrometria de massas (LC-MS/MS) para quantificação do lorazepam em plasma humano visando aplicação em estudo de bioequivalência entre duas formulações de comprimidos contendo esse fármaco. A preparação das amostras de plasma foi feita por extração líquido-líquido usando hexano:diclorometano (60:40 v/v) como solvente de extração. O padrão interno usado foi o bromazepam. A separação cromatográfica ocorreu utilizando-se coluna analítica modelo Gemini® C18 110 A (150 mm x 4,6 mm; partículas de 5µm). Mistura de metanol e tampão acetato de amônio 10 mM (80:20, v/v), acrescida de 0,1% de ácido fórmico ao final da preparação, foi usada como fase móvel. A interface entre HPLC e MS/MS foi a fonte de ionização por eletrospray (ESI) operando em modo positivo (ES+). O analito e o PI foram monitorados e quantificados através de multiple reaction monitoring (MRM). As transições monitoradas foram m/z 320,69 > 274,96 para o lorazepam e m/z 318,00 > 182,20 para o padrão interno. O método foi validado na faixa de concentração de 0,50 a 80,0 ng/ml em plasma humano. A bioequivalência entre as formulações foi determinada através dos intervalos de confiança 90 % obtidos para as razões dos parâmetros farmacocinéticos Cmax (99% - 114%), AUC0-t (93% - 105%) e AUC0-inf (96% - 107%). Concluiu-se que as duas formulações podem ser administradas de maneira intercambiável sem prejuízo da eficácia terapêutica.
Título en inglés
A bioanalytical method using liquid chromatography coupled to MS/MS detection system for the quantification of Lorazepam in human plasma aiming bioequivalence studies.
Palabras clave en inglés
Bioequivalence
Bromazepam
Human plasma.
LC-MS/MS
Lorazepam
Resumen en inglés
A method of liquid chromatography coupled to mass spectrometric detection (LC-MS/MS) with high sensitivity and specificity was developed to quantify Lorazepam in human plasma. This method was applied in a bioequivalence study between two tablet formulations. The preparation of plasma samples were performed by liquid-liquid extraction using hexanedichloromethane (60:40 v/v) as extraction solvent. The internal standard was bromoazepam. The chromatographic separation was achieved using the Gemini® C18 110A (150 mm x 4.6mm; 5µm particles) analytical column. The mobile phase was prepared from a mixture of methanol and 10mM of ammonium acetate (80:20, v/v), and finally adding 0.1% of formic acid. The HPLC and MS/MS interface was the electrospray ionization source (ESI), operating in positive mode (ESI+). The analyte and internal standard were monitored and quantified through multiple reaction monitoring (MRM). The monitored transitions were m/z 320.69 > 274.96 for lorazepam and m/z 318.00 > 182.20 for the internal standard. The method was validated over the range 0.50 to 80.0 ng/mL in human plasma. The bioequivalence between the two formulations was determined inside the 90% confidence interval for the pharmacokinetic parameters, Cmax (99% - 114%), AUC0-t (93% - 105%) and AUC0-inf (96% - 107%). It was concluded that the two formulations can be administered in an interchangeable manner without losing the therapeutic efficiency.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2013-07-17
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.