Master's Dissertation
DOI
https://doi.org/10.11606/D.9.2009.tde-01022010-161804
Document
Author
Full name
Glaucia Cristina Mello Santos
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2009
Supervisor
Committee
Pinto, Terezinha de Jesus Andreoli (President)
Fonseca, Maria Jose Vieira
Salgado, Hérida Regina Nunes
Title in Portuguese
Redução da biocarga e garantia de esterilidade em implantes mamários de silicone
Keywords in Portuguese
Biocarga
Biomateriais
Calor seco
Esterilização
Implante mamário
Silicone
Vulcanização
Abstract in Portuguese
Os implantes mamários de silicone constituem-se em biomateriais os quais têm sido amplamente utilizados em cirurgias para reconstituição da mama após ocorrência de câncer, acidentes, correção do tamanho dos seios quando há uma diferença de volume entre eles, ou até mesmo, para o aumento do tamanho da mama por motivos estéticos. Com o intuito de conferir segurança em sua aplicação, intimamente ligada à saúde dos pacientes, grande atenção tem sido dada aos processos de esterilização aplicados a biomateriais. Ao avaliar o processo produtivo de implantes mamários de silicone, um dos processos de esterilização mais comumente empregados é o calor seco, que necessita de elevadas temperaturas por longo tempo para o sucesso da esterilização. A exposição do implante a tais condições potencialmente pode ocasionar alterações de suas características. De outro lado, uma etapa preliminar do processo produtivo do implante é a vulcanização, que consiste no aquecimento do implante a temperaturas da ordem de 165 ± 5°C por aproximadamente 9 horas. Considerando tempo e temperatura empregados nesta etapa, o objetivo deste estudo foi avaliar a carga microbiana dos implantes mamários de silicone antes do processo de vulcanização, assim como o decaimento da carga microbiana neste processo e também confirmar a esterilidade do gel contido internamente à membrana. Desta forma, foi possível observar que o nível de contaminação microbiana dos implantes gelatinosos é relativamente baixo e que a vulcanização foi um processo que possibilitou a inativação de até 108 esporos, a concentração de esporos mais alta utilizada no estudo. Os resultados mostraram que a vulcanização possibilitou não só a redução da carga microbiana, mas também consiste em mecanismo para garantir a esterilidade do gel interno ao produto. Desta forma, o processo esterilizante final teve como contribuição elevar o Nível de Garantia de Esterilidade (SAL ou Sterility Assurance Level), condição interessante ao se considerar a tendência de adoção da liberação paramétrica, assim como o conceito de validação combinado bioburden/indicador biológico em vez de sobre-morte. Avaliação complementar foi feita ao quantificar endotoxina nos implantes antes e após o processo de esterilização (calor seco e óxido de etileno), verificando-se que os processos considerados não alteram significativamente a quantidade de endotoxina. Ainda assim, em todas as situações foram obtidos níveis aceitáveis, conforme USP 31.
Title in English
Bioburden reduction and sterility assurance in silicone gel breast implants
Keywords in English
Bioburden
Biomaterials
Breast implant
Dry heat
Silicone
Sterilization
Vulcanization
Abstract in English
Silicone breast implants consist of biomaterials widely used in breast reconstitution surgeries after the occurrence of cancer, accidents, breast size correction (in case of different volume between both breasts)or in mammary augmentation for esthetic reasons. With a view to confer security in its application, directly related to patients health, great attention has been given to sterilization processes applied to biomaterials. Among these, dry heat is one of the most often employed. For a successful sterilization, it requires high temperatures for a long period, what may give rise to alteration of the implant characteristics. On the other hand, a preliminary stage of the implant production process is vulcanization, which consists of heating the implant to 165 ± 5°C for approximately 9 hours. Taking into account the time and temperature used in this stage, the aim of this work was to evaluate the bioburden of silicone breast implants prior to the vulcanization process and the decline in bioburden due to this process, and to confirm the sterility of the gel contained in the membrane. This study led us to the conclusion that the level of microbial contamination of gel implants is relatively low, and that vulcanization allowed for the inactivation of up to 100 million spores, the highest concentration of spores used in this study. The results obtained showed that vulcanization enabled not only the reduction of the microbial load, but also guaranteed the sterility of the gel inside the product. Thus, the final sterilizing process contributed to an increase in the Sterility Assurance Level, an interesting phenomenon if we consider the tendency toward adoption of parametric release and the concept of a combined validation bioburden/biological indicator rather than overkill. Complement evaluation was made measuring endotoxins in the implants before and after the sterilization process (dry heat and ethylene oxide), verifying that the considered processes do not modify the amount of endotoxin significantly, as expectation. Still thus, in all the situations had been gotten acceptable levels, as USP 31.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2010-03-11
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.