Titre en portugais
Implantes porosos à base de titânio, avaliação in vitro e in vivo
Mots-clés en portugais
liga Ti-13Nb-13Zr
osteointegração
porosidade
titânio
Resumé en portugais
O desenvolvimento de biomateriais com estrutura porosa visa permitir uma boa ancoragem biológica através do crescimento do tecido ósseo no interior dos poros, além de no caso de algumas ligas de titânio, proporcionar valores de módulo elástico semelhantes ao do osso, prevenindo assim o insucesso por reabsorção óssea na interface com o implante. O objetivo deste estudo foi avaliar o processamento de implantes porosos à base de titânio produzidos por metalurgia do pó, utilizando polímeros naturais como aditivos (albumina, amidos de arroz, batata, milho e a gelatina), e analisar a osteointegração desses implantes por ensaio in vivo. Foram obtidas, amostras de titânio comercialmente puro (Ticp) e da Liga Ti-13Nb-13Zr pelos processos de space-holder e suspensão; os polímeros naturais foram removidos termicamente (350°C/1h) previamente a sinterização (1300°C/1h e 1300°C/3h, respectivamente). As metodologias de processamento utilizando aditivos de baixo custo propiciaram a obtenção de implantes metálicos porosos de maneira simplificada, com porosidades diferenciadas, boa homogeneidade estrutural, grau de porosidade adequado (entre 40% e 60%), distribuição e tamanho dos poros conforme o polímero natural formador. Na caracterização biológica, o teste in vitro de citotoxicidade validou a utilização dos implantes para a realização do teste in vivo. A avaliação da osteointegração foi realizada em coelhos da raça New Zealand em períodos de 49 dias. Na análise histológica por microscopia eletrônica de varredura e fluorescência, foram obtidos os resultados quantitativos e qualitativos da osteointegração, com crescimento ósseo em todos os implantes, apresentando osteocondutividade. Os melhores resultados ocorreram em implantes com porosidade homogênea, e com tamanho médio de poros entre 100 e 200 μm. Assim, os materiais porosos de titânio e da liga Ti-13Nb-13Zr produzidos com os polímeros naturais (amido de arroz, amido de batata, amido de milho, gelatina e albumina), permitiram a nutrição e manutenção necessárias para a sobrevivência das células, comprovando a osteointegração e osteocondução dos implantes desenvolvidos.
Titre en anglais
Titanium based porous implants, in vitro and in vivo evaluation
Mots-clés en anglais
osseointegration
porosity
Ti-13Nb-13Zr alloy
titanium
Resumé en anglais
The development of biomaterials with porous structure, seeks to allow a good biological anchorage through bone tissue ingrowth within the pores, besides to propitiate elastic modulus values similar to bone, thus preventing the failure by bone resorption, on implants interface. The goals of this study was to evaluate the processing of titanium based porous implants by powder metallurgy (PM), using natural polymers as additive (rice, potato and corn starches, albumin and gelatin), and evaluate these implants osseointegration by in vivo assay. Samples were obtained of commercially pure titanium (cpTi) and Ti-13Nb-13Zr alloy by space-holder and suspension techniques; the natural polymers were removed by thermal treatment (350C/1h) prior to sintering (1300C/1h and 1300C/3h respectively). The processing methodologies using low cost additives propitiate the production of porous metallic implants in a simplified manner, with differentiated porosities, good structural homogeneity, proper porosity degree (between 40% and 60%), distribution and pore size as the related natural polymer. The biological characterization was first performed in vitro with a cytotoxicity test, which allowed the use of implants in vivo. The osseointegration evaluation was performed in New Zealand White rabbits for 49 days period. Histological analyses with scanning electron microscopy and fluorescence microscopy were obtained quantitative and qualitative results of osseointegration, with bone growth in all implants, presenting osteoconductivity. The best results were in implants with homogenous porosity and with mean pore size between 100 and 200 μm. Thus, titanium and Ti-13Nb-13Zr alloy porous materials produced with natural polymers (rice, potato and corn starches, albumin and gelatin), allowed the nutrition and nourishing needed to bone cells survival, proving the osseointegration and osteoconduction features of developed implants.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
2013GoiaImplantes.pdf (12.22 Mbytes)
Date de Publication
2013-06-06
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.