Tesis Doctoral
DOI
https://doi.org/10.11606/T.85.2013.tde-29052013-094946
Documento
Autor
Nombre completo
Tamiye Simone Goia
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2013
Director
Tribunal
Bressiani, Ana Helena de Almeida (Presidente)
Allegrini Junior, Sergio
Higa, Olga Zazuco
Takiishi, Hidetoshi
Zavaglia, Cecília Amélia de Carvalho
Título en portugués
Implantes porosos à base de titânio, avaliação in vitro e in vivo
Palabras clave en portugués
liga Ti-13Nb-13Zr
osteointegração
porosidade
titânio
Resumen en portugués
O desenvolvimento de biomateriais com estrutura porosa visa permitir uma boa ancoragem biológica através do crescimento do tecido ósseo no interior dos poros, além de no caso de algumas ligas de titânio, proporcionar valores de módulo elástico semelhantes ao do osso, prevenindo assim o insucesso por reabsorção óssea na interface com o implante. O objetivo deste estudo foi avaliar o processamento de implantes porosos à base de titânio produzidos por metalurgia do pó, utilizando polímeros naturais como aditivos (albumina, amidos de arroz, batata, milho e a gelatina), e analisar a osteointegração desses implantes por ensaio in vivo. Foram obtidas, amostras de titânio comercialmente puro (Ticp) e da Liga Ti-13Nb-13Zr pelos processos de space-holder e suspensão; os polímeros naturais foram removidos termicamente (350°C/1h) previamente a sinterização (1300°C/1h e 1300°C/3h, respectivamente). As metodologias de processamento utilizando aditivos de baixo custo propiciaram a obtenção de implantes metálicos porosos de maneira simplificada, com porosidades diferenciadas, boa homogeneidade estrutural, grau de porosidade adequado (entre 40% e 60%), distribuição e tamanho dos poros conforme o polímero natural formador. Na caracterização biológica, o teste in vitro de citotoxicidade validou a utilização dos implantes para a realização do teste in vivo. A avaliação da osteointegração foi realizada em coelhos da raça New Zealand em períodos de 49 dias. Na análise histológica por microscopia eletrônica de varredura e fluorescência, foram obtidos os resultados quantitativos e qualitativos da osteointegração, com crescimento ósseo em todos os implantes, apresentando osteocondutividade. Os melhores resultados ocorreram em implantes com porosidade homogênea, e com tamanho médio de poros entre 100 e 200 μm. Assim, os materiais porosos de titânio e da liga Ti-13Nb-13Zr produzidos com os polímeros naturais (amido de arroz, amido de batata, amido de milho, gelatina e albumina), permitiram a nutrição e manutenção necessárias para a sobrevivência das células, comprovando a osteointegração e osteocondução dos implantes desenvolvidos.
Título en inglés
Titanium based porous implants, in vitro and in vivo evaluation
Palabras clave en inglés
osseointegration
porosity
Ti-13Nb-13Zr alloy
titanium
Resumen en inglés
The development of biomaterials with porous structure, seeks to allow a good biological anchorage through bone tissue ingrowth within the pores, besides to propitiate elastic modulus values similar to bone, thus preventing the failure by bone resorption, on implants interface. The goals of this study was to evaluate the processing of titanium based porous implants by powder metallurgy (PM), using natural polymers as additive (rice, potato and corn starches, albumin and gelatin), and evaluate these implants osseointegration by in vivo assay. Samples were obtained of commercially pure titanium (cpTi) and Ti-13Nb-13Zr alloy by space-holder and suspension techniques; the natural polymers were removed by thermal treatment (350C/1h) prior to sintering (1300C/1h and 1300C/3h respectively). The processing methodologies using low cost additives propitiate the production of porous metallic implants in a simplified manner, with differentiated porosities, good structural homogeneity, proper porosity degree (between 40% and 60%), distribution and pore size as the related natural polymer. The biological characterization was first performed in vitro with a cytotoxicity test, which allowed the use of implants in vivo. The osseointegration evaluation was performed in New Zealand White rabbits for 49 days period. Histological analyses with scanning electron microscopy and fluorescence microscopy were obtained quantitative and qualitative results of osseointegration, with bone growth in all implants, presenting osteoconductivity. The best results were in implants with homogenous porosity and with mean pore size between 100 and 200 μm. Thus, titanium and Ti-13Nb-13Zr alloy porous materials produced with natural polymers (rice, potato and corn starches, albumin and gelatin), allowed the nutrition and nourishing needed to bone cells survival, proving the osseointegration and osteoconduction features of developed implants.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
2013GoiaImplantes.pdf (12.22 Mbytes)
Fecha de Publicación
2013-06-06
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.