Doctoral Thesis
DOI
https://doi.org/10.11606/T.85.2009.tde-22092011-100341
Document
Author
Full name
Jose Pucci Caly
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2009
Supervisor
Committee
Nogueira, Gessé Eduardo Calvo (President)
Freitas, Anderson Zanardi de
Landulfo, Eduardo
Muramatsu, Mikiya
Teixeira, Julio Carlos
Title in Portuguese
Estudo e avaliação da radiometria no tratamento fototerápico da hiperbilirrubinemia neonatal
Keywords in Portuguese
fototerapia
radiometria
radiometro
responsividade espacial
responsividade espectral
Abstract in Portuguese
A fototerapia é um procedimento estabelecido há mais de 50 anos no tratamento da icterícia em recém-nascidos. No entanto ainda não há um método padronizado de quantificação da dose fototerápica nos estudos clínicos publicados, dificultando a comparação de estudos prévios sobre a eficácia da fototerapia, bem como o estabelecimento de doses seguras e previsíveis. A dose fototerápica depende, dentre outros fatores, da irradiância média útil entregue pelas unidades fototerápicas. Mas não há procedimentos padronizados tanto na quantificação da irradiância útil ao tratamento, quanto na estimação da irradiância média útil em fontes fototerápicas. Por isso, têm sido relatadas grandes variações nas medições de uma mesma fonte usando os radiômetros disponíveis no mercado, como conseqüência da vasta diversidade de responsividades espectrais dos instrumentos. Um objetivo deste trabalho foi adaptar e aplicar os princípios da radiometria ultravioleta de banda larga na quantificação da irradiância disponível em fontes fototerápicas, estabelecendo procedimentos que permitem comparar irradiâncias medidas em diversas fontes, usando radiômetros com diferentes responsividades espectrais. Outro objetivo foi caracterizar amostras de fontes comumente usadas atualmente, focando o problema da estimação da irradiância média útil em unidades de fototerapia, propondo um método de estimação da irradiância média útil em fontes com foco dirigido. Os resultados experimentais obtidos nos permitem concluir que não somente é necessário padronizar a radiometria fototerápica, como também o método de estimação da irradiância média útil.
Title in English
Study and evaluation of radiometry in phototherapeutic treatment of the neonatal hiperbilirubinemia
Keywords in English
phototherapy
radiometry
spectral responsivities
Abstract in English
Phototherapy is a procedure established more than 50 years ago in the treatment of the newborn jaundice. However there is no a standard method to quantify the phototherapeutic dose in published clinical studies, hindering the comparison of previous studies on phototherapeutic effectiveness, as well as the establishment of safe and predictable doses. The phototherapeutic dose depends, among other factors, on the effective mean irradiance produced by the phototherapeutic unit. There are no standard procedures, however, neither to quantify the effective irradiance, nor to estimate the mean effective irradiance. As a consequence, large measurement variations in a same phototherapeutic unit are observed using different commercially available radiometers, as a consequence of the vast diversity of spectral responsivities of the instruments. An objective of this work was to adapt and to apply the bases of the wideband ultraviolet radiometry to quantify the available irradiance from phototherapeutic units, establishing procedures that allow us to compare measured irradiances from different sources, using radiometers presenting different spectral responsivities. Another objective was to characterize samples of phototherapeutic units commonly used, focusing the problem of the estimation of the effective mean irradiance from phototherapeutic units, proposing a method to estimate of the effective irradiance from focused sources. The experimental results allow us to conclude that it is not only necessary to standardize the phototherapeutic radiometry, but also the method of estimation of the effective mean irradiance.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
2009CalyEstudo.pdf (640.38 Kbytes)
Publishing Date
2011-10-17
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.