Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2009.tde-29062009-153847
Documento
Autor
Nombre completo
Osvaldo Luiz da Costa
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2009
Director
Tribunal
Sciani, Valdir (Presidente)
Santos, Gonçalo Rodrigues dos
Sumiya, Luiz Carlos do Amaral
Título en portugués
Otimização do sistema de carga e descarga no porta-alvo de irradiação  de H218O utilizando a produção de 18F- no Cíclotron Cyclone 30 do IPEN-CNEN/SP
Palabras clave en portugués
cíclotron
FDG
flúor
PET
Resumen en portugués
O aumento da demanda no Brasil pelo radiofármaco [18F]FDG para exames de tomografia por emissão de pósitron (PET-CT) e sua meia-vida de 109,7 minutos exigem uma atenção especial para a cadeia produtiva deste radiofármaco. Desde a irradiação da água enriquecida em 18O até a passagem do paciente pelo tomógrafo, numa seqüência de procedimentos que pode durar até cerca de seis horas, todas as etapas da cadeia produtiva devem ter a maior confiabilidade possível, pois qualquer falha em uma das etapas será sentida na extremidade da cadeia produtiva. A falta de indicação de posição da válvula do Sistema de Carga e Descarga do Porta-alvo de 18F- no Centro de Aceleradores Cíclotron levou a perda de produção de 18F-, contaminação da Sala de Irradiação e aumento da dose dos trabalhadores envolvidos na operação e manutenção dos sistemas de irradiação. Este trabalho testou o comportamento de três tipos de sensores de posição (microswitch, reedswitch e sensor de proximidade indutivo) no ambiente da Sala de Irradiação 1.2 do Centro de Aceleradores Cíclotron, que está sujeito a altas taxas de radiação gama e nêutrons devido à produção rotineira de 18F- e 123I, através deste teste foi possível encontrar o sensor de posição mais adequado para trabalhar no porta-alvo de 18F-, e através da reprogramação do controlador lógico programável, foi possível evitar este tipo de falha durante a produção de 18F- no Centro de Aceleradores Cíclotron, e aumentar a confiabilidade na cadeia produtiva do [18F]FDG.
Título en inglés
LOAD AND UNLOAD SYSTEM OPTIMIZATION ON H2 18O IRRADIATION TARGET USED FOR 18F- PRODUCTION AT THE CYCLOTRON CYCLONE 30 FROM IPEN-CNEN/SP
Palabras clave en inglés
cíclotron
FDG
flúor
PET
Resumen en inglés
The demand growing in Brazil by the radiopharmaceutical [18F]FDG in positron emission tomography (PET-CT) and the 109,7 minutes half life claim special attention to the productive chain of this radiopharmaceutical. Since the [18O]water irradiation until the tomograph patient scanning, in sequential procedures that may spent about six hours, all the productive chain stages must be as reliable as possible, because any stage failed will be perceived in productive chain extremity. The position indication absence from Load and Unload 18F- Target System valve in Cyclotron Accelerators Center resulted in 18F- production loss, Irradiation Room contamination and the increase workers dose responsible by operation and maintenance of irradiation systems. This study tested the behaviour of three types of position sensors (microswitch, reedswitch and inductive sensor), into Irradiation Room 1.2 environment of the Cyclotron Accelerators Center, where there are high gamma radiation and neutrons rates because the routine 18F- and 123I production, through this test was possible to discover the fitter position sensor to run on 18F- Target, and after rewriting the programmable logic controller software was possible avoid this type of fail at 18F- production time in Cyclotron Accelerators Center, and to grow up the reliability on [18F]FDG productive chain.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2009-07-16
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.