Doctoral Thesis
DOI
https://doi.org/10.11606/T.85.2007.tde-27112007-144824
Document
Author
Full name
Maria Esmeralda Ramos Poli
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2007
Supervisor
Committee
Rodrigues, Letícia Lucente Campos (President)
Furquim, Tânia Aparecida Correia
Ghilardi Netto, Thomaz
Rocha, Felícia Del Gallo
Rodrigues, Laura Natal
Title in Portuguese
Definição do volume de planejamento do alvo (PTV) e seu efeito na radioterapia
Keywords in Portuguese
Movimento de Órgãos
PTV
Radioterapia
Volume alvo
Abstract in Portuguese
Este trabalho visa estudar as margens necessárias para definir o volume de planejamento do alvo (PTV) requeridas para tratar adequadamente tumores móveis como os localizados na próstata ou tumores localizados em áreas com pouca mobilidade como os da região da cabeça e pescoço, na ausência de localização do alvo por imagem. Também tem como objetivo avaliar o impacto causado pelo PTV, em termos de dose, nas estruturas críticas ao seu redor e sua influência quando planejamento inverso é utilizado na radioterapia com modulação de feixe (IMRT). Dados de 387 pacientes de próstata foram analizados retrospectivamente. Todos os pacientes receberam localização pré-tratamento com ultra-som 2D resultando em 10.327 localizações, cada uma com deslocamento de isocentro em três direções: antero-posterior (AP), lateral direitaesquerda (DE), e superior-inferior (SI). O deslocamento médio e desvio padrão (SD) para cada direção foi estimado a partir dos dados de tratamento gravados diariamente. As incertezas (SD) na posição do alvo foram 4,4 mm (AP), 3,6 mm (DE), e 4,5 mm (SI). O estudo das incertezas no posicionamento diário de 78 pacientes com tumores de cabeça e pescoço que utilizaram máscaras termoplásticas como imobilizadores, avaliados com equipamento de portal com imagem eletrônica (EPID), mostrou variações (SD) na posição do isocentro de tratamento de 3,1 mm (AP), 1,5 mm (DE), 4,5 mm (SI). Aplicando estes desvios num simulador antropomórfico estudou-se os histogramas de dose-volume resultantes do deslocamento do isocentro no tratamento diário. Os resultados mostraram a importância de se colocar margens no volume clínico do alvo para garantir um tratamento adequado e também mostraram que a variação diária do isocentro de tratamento pode causar um aumento de dose maior que o nível de tolerância dos órgãos críticos.
Title in English
Planning target volume (PTV) definition and its effects in the radiotherapy
Keywords in English
Organs mobile
PTV
Radiotherapy
Target volume
Abstract in English
This work intends to study the margins required to define a planning target volume (PTV) for adequate treatment of the mobile tumors such as prostate or those located in areas with less mobility as the ones in head and neck region, in the absence of daily localization imaging based. It is also intends to evaluate the impact caused by the PTV, in terms of dose, to the critical structures surrounding the PTV and its influence when inverse planning is used in the intensity-modulated radiation therapy (IMRT). Data from 387 prostate patients were analyzed retrospectively. Every patient in the study received daily pre-treatment localization with 2D ultrasound resulting in a total of 10,327 localizations, each comprising of an isocenter displacement in 3 directions: anterior-posterior (AP), right-left lateral (RL), and superior-inferior (SI). The mean displacement and standard deviation (SD) for each direction for each patient was computed from daily treatment records. The uncertainties (SD) in the target position were 4.4 mm (AP), 3.6 mm (RL), and 4.5 mm (SI). A study of the uncertainties in the daily positioning of 78 head and neck patients who used thermoplastic mask to immobilize them, evaluated with electronic portal imaging device (EPID), showed variations (SD) in the isocenter treatment position of 3.1 mm (AP), 1.5 mm (RL), and 4.5 mm (SI). By applying these shifts in an anthropomorphic phantom it was studied the dose-volume histograms resultant of the isocenter displacement in the daily treatment. The result showed the importance of putting margins in the clinical target volume to assure an adequate treatment and also showed that isocenter daily variation can cause an increase to the dose greater than the tolerance level to the critical organs.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2007-12-10
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors