Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2006.tde-27032012-084827
Documento
Autor
Nombre completo
Patricia Mára de Siqueira
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2006
Director
Tribunal
Rodrigues, Laura Natal (Presidente)
Todo, Alberto Saburo
Yoshimura, Elisabeth Mateus
Título en portugués
Implantação do novo protocolo de dosimetria da AIEA no LCI/IPEN/CNEN
Palabras clave en portugués
dose absorvida
laboratórios de dosimetria
protocolos de dosimetria
Resumen en portugués
A fim de implementar o novo código de prática da Agência Internacional de Energia Atômica (AIEA) no Brasil, o Laboratório de Calibração de Instrumentos (LCI/IPEN) está realizando calibrações em termos de kerma no ar e em termos de dose absorvida na água, em feixes de 60Co. As razões ND,w/NK obtidas são comparadas com valores da literatura, obtendo-se concordância satisfatória. As diferenças entre os valores das razões CK (ND,w/NK) obtidas no presente trabalho e os valores da literatura são devidas a vários fatores. Estes fatores podem ser as variações entre as câmaras, que devem ser objeto de diferenças nas incertezas estimadas pelos PSDLs, e as diferenças nos padrões utilizados por cada Instituto de Metrologia Nacional (NMI) ou Laboratório de Dosimetria. No entanto, se forem conhecidas as razões entre cada NMI e o BIPM, para os padrões de kerma no ar e de dose absorvida na água, é possível a normalização das razões CK medidas para as razões equivalentes ao BIPM. Todos os resultados de razões CK obtidos foram convertidos para as razões equivalentes ao BIPM para facilitar a comparação. Neste trabalho é discutida a utilização da razão CK como parâmetro de controle de qualidade na verificação de resultados das calibrações rotineiras. Para avaliação dos procedimentos de calibração adotados no LCI, foram realizados testes de estabilidade a longo prazo com as câmaras de ionização padrões. Além disso, foi criado um banco de dados para armazenamento dos dados de calibração, com o objetivo de melhorar a qualidade do procedimento de calibração.
Título en inglés
Implementation of the new IAEA code of practice at LCI/IPEN/CNEN
Palabras clave en inglés
calibration standards
dosimetry
iaea
ionization chambers
kerma
laboratories
radiation doses
recommendations
Resumen en inglés
In order to implement the new IAEA code of practice in Brazil the national calibration laboratories, the Laboratory of Instrument Calibration (LCI/IPEN) is calibrating clinical dosimeters in terms of both air kerma and absorbed dose to water in a 60Co gamma ray beam. The ND,w/NK ratios thus obtained are then compared with the literature values; a satisfactory agreement has been found. The differences between the CK values obtained in the present work and the literature values may be due to several components. These could be the chamber-to-chamber variations which may be subject to different estimated uncertainties by PSDLs, and the variations in the standards used by each National Metrology Institute (NMI) or Dosimetry Laboratory. However, since the ratio of both air kerma and absorbed dose to water standards for each NMI to those of the BIPM are known, it is possible to reduce the measured ratios to the expected BIPM equivalent ratios. All these CK results have been converted to the BIPM equivalent values to facilitate comparison. The use of the CK value as a and quality control parameter to verify the results of the routine calibrations is discussed in this work. The long term stability of wich standard ionization chamber was checked in order to evaluate the calibration procedures followed at LCI. Moreover, a database was created to store the calibration data aiming the quality improvement of the calibration procedure.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2012-04-04
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.