Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.85.2012.tde-24082012-162508
Document
Auteur
Nom complet
Roberto Manuel dos Santos
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2012
Directeur
Jury
Genezini, Frederico Antonio (Président)
Maidana, Nora Lia
Zahn, Guilherme Soares
Titre en portugais
Desenvolvimento de um método para obtenção da eficiência de detecção para detectores HPGe em medidas com fontes extensas
Mots-clés en portugais
calibração de eficiência
fonte extensa
média geométrica
Resumé en portugais
Um dos grandes desafios em medidas de espectrometria gama utilizando fontes que não podem ser consideradas pontuais é a determinação da eficiência de detecção do sistema, uma vez que a geometria da fonte impõe complexidades na tarefa como o tratamento do ângulo sólido e da auto-absorção. Alguns métodos para o cálculo dessa eficiência já são bem estabelecidos e podem ser encontrados na literatura. Neste trabalho é proposto um método com características distintas aos já existentes cujas principais vantagens são, não depender de conhecimento prévio da geometria de detecção e da composição química da amostra a ser medida, e permitir o cálculo da eficiência de uma fonte extensa a partir de medidas com fontes padrão em formato pontual. O método proposto foi testado comparando os valores das eficiências obtidas por ele com os valores obtidos usando uma fonte padrão extensa. Para testar o método proposto em diferentes condições foi utilizado solo como matriz para a amostra em duas diferentes espessuras, 1cm e 3cm. Testou-se também o comportamento do método quando a matriz tem número atômico efetivo mais alto (aqui utilizou-se solo misturado à limalha de ferro) e mais baixo (adição de pó de grafite ao solo). Os resultados indicaram que o método é adequado para amostras de solo com espessura mais fina; já para fontes mais espessas, dois dos três resultados indicaram que a eficiência foi subestimada pelo método proposto, sugerindo que uma geometria mais adequada deva ser escolhida para a amostragem das medidas
Titre en anglais
Development of a method for obtaining the efficiency of detection in hpge detector with measures large sources
Mots-clés en anglais
efficiency calibration
geometric average
large sources
Resumé en anglais
One of the greatest challenges in gamma spectroscopy measurements using extense sources is the determination of the system's detection efficiency, since the source's geometry leads to added complexity to the treatment of parameters as geometry and self-absorption. Some well-established methods for the efficiency determination can be found in the literature. In the present work a method with distinct characteristics is proposed with the main advantages that it doesn't require previous knowledge of the detection geometry or of the sample composition; moreover, it allows the determination of the detection efficiency for a large sample based only on measurements using a point-like standard source. The proposed method was tested comparing the efficiency values obtained using the method with the values obtained using a real large source with the same composition and geometry. In order to verify the performance of the method under different conditions, soil matrixes of two different thicknesses (1cm and 3cm) were used; also, to asses the method under different sample compositions, 3cm-thick soil matrixes with increased effective atomic number (where iron filings was added) and with reduced effective atomic mass (where graphite powder was added) were used. The results indicate that the proposed method is suitable for thinner soil samples; as for thicker samples, two of the three results showed a subestimation of the detection efficiency value determined, suggesting that a more suitable sampling geometry should be analyzed.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-11-09
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.