Master's Dissertation
DOI
https://doi.org/10.11606/D.85.2012.tde-24082012-162508
Document
Author
Full name
Roberto Manuel dos Santos
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2012
Supervisor
Committee
Genezini, Frederico Antonio (President)
Maidana, Nora Lia
Zahn, Guilherme Soares
Title in Portuguese
Desenvolvimento de um método para obtenção da eficiência de detecção para detectores HPGe em medidas com fontes extensas
Keywords in Portuguese
calibração de eficiência
fonte extensa
média geométrica
Abstract in Portuguese
Um dos grandes desafios em medidas de espectrometria gama utilizando fontes que não podem ser consideradas pontuais é a determinação da eficiência de detecção do sistema, uma vez que a geometria da fonte impõe complexidades na tarefa como o tratamento do ângulo sólido e da auto-absorção. Alguns métodos para o cálculo dessa eficiência já são bem estabelecidos e podem ser encontrados na literatura. Neste trabalho é proposto um método com características distintas aos já existentes cujas principais vantagens são, não depender de conhecimento prévio da geometria de detecção e da composição química da amostra a ser medida, e permitir o cálculo da eficiência de uma fonte extensa a partir de medidas com fontes padrão em formato pontual. O método proposto foi testado comparando os valores das eficiências obtidas por ele com os valores obtidos usando uma fonte padrão extensa. Para testar o método proposto em diferentes condições foi utilizado solo como matriz para a amostra em duas diferentes espessuras, 1cm e 3cm. Testou-se também o comportamento do método quando a matriz tem número atômico efetivo mais alto (aqui utilizou-se solo misturado à limalha de ferro) e mais baixo (adição de pó de grafite ao solo). Os resultados indicaram que o método é adequado para amostras de solo com espessura mais fina; já para fontes mais espessas, dois dos três resultados indicaram que a eficiência foi subestimada pelo método proposto, sugerindo que uma geometria mais adequada deva ser escolhida para a amostragem das medidas
Title in English
Development of a method for obtaining the efficiency of detection in hpge detector with measures large sources
Keywords in English
efficiency calibration
geometric average
large sources
Abstract in English
One of the greatest challenges in gamma spectroscopy measurements using extense sources is the determination of the system's detection efficiency, since the source's geometry leads to added complexity to the treatment of parameters as geometry and self-absorption. Some well-established methods for the efficiency determination can be found in the literature. In the present work a method with distinct characteristics is proposed with the main advantages that it doesn't require previous knowledge of the detection geometry or of the sample composition; moreover, it allows the determination of the detection efficiency for a large sample based only on measurements using a point-like standard source. The proposed method was tested comparing the efficiency values obtained using the method with the values obtained using a real large source with the same composition and geometry. In order to verify the performance of the method under different conditions, soil matrixes of two different thicknesses (1cm and 3cm) were used; also, to asses the method under different sample compositions, 3cm-thick soil matrixes with increased effective atomic number (where iron filings was added) and with reduced effective atomic mass (where graphite powder was added) were used. The results indicate that the proposed method is suitable for thinner soil samples; as for thicker samples, two of the three results showed a subestimation of the detection efficiency value determined, suggesting that a more suitable sampling geometry should be analyzed.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2012-11-09
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.