Disertación de Maestría
DOI
https://doi.org/10.11606/D.85.2001.tde-23072002-091706
Documento
Autor
Nombre completo
Patricia Lammoglia
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2001
Director
Tribunal
Caldas, Linda Viola Ehlin (Presidente)
Medeiros, Regina Bitelli
Terini, Ricardo Andrade
Título en portugués
"Elaboração e Implementação de Testes de Controle de Qualidade em Equipamentos de Angiografia por Subtração Digital"
Palabras clave en portugués
Angiografia
Controle de Qualidade
Fluoroscopia
Garantia de Qualidade
Resumen en portugués
Foram elaborados e implementados testes de controle de qualidade em equipamentos de angiografia por subtração digital. Estes testes foram baseados em normas nacionais e internacionais e foram implementados em cinco equipamentos de três instituições diferentes. Estes testes avaliam os parâmetros de desempenho dos equipamentos e são utilizados para a determinação da taxa de kerma no ar na entrada da pele do paciente e do corpo clínico. Os testes foram realizados utilizando dispositivos, simuladores de pacientes e câmaras de ionização. Os testes convencionais e do sistema de subtração digital indicaram que os equipamentos avaliados encontravam-se em bom estado de desempenho. Os testes para verificação da taxa de kerma no ar indicaram que um dos equipamentos avaliados apresentou altas taxas de kerma no ar na entrada da pele do paciente em modo fluoroscopia. Este problema foi notificado, e as devidas providências foram imediatamente tomadas pela gerência do hospital.
Título en inglés
Elaboration and Implementation of Quality Control Tests in Digital Subtraction Angiography Equipments
Palabras clave en inglés
Angiography
fluoroscopy
Quality Control
Resumen en inglés
Quality control tests were elaborated and implementated in digital subtraction angiography equipments. These tests were elaborated based in national and international standards, and they were implementated in five equipments of three different institutions. These tests are utilized to evaluate the performance of these equipments and to determine the patient entrance and personnel air kerma rates. The tests were performed using test tools, patient phantoms e ionization chambers. The conventional tests and the tests of the digital subtraction systems indicated that the evaluated equipments presented a good performance. The tests implementated to determine the air kerma rates presented, in one case (fluoroscopy equipment), high patient entrance skin air kerma rate. This problem was notified, and the providences were promptly taken by the hospital administration.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2004-07-16
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.